In vitro investigation of two connector types for continuous rod construct to extend lumbar spinal instrumentation
Bastian Welke1, Michael Schwarze2, Christof Hurschler2
1Laboratory for Biomechanics and Biomaterials, Department of Orthopaedic Surgery, Hannover Medical School, Anna-von-Borries-Str. 1-7, 30625, Hannover, Germany. welke.bastian@mh-hannover.de.
Purpose:
Instrumentation of the lumbar spine is a common procedure for treating pathologic conditions. Studies have revealed the risks of pathologies in the adjacent segments, with the incidence rate being up to 36.1%. Revision procedures are often required, including extension of the instrumentation by the use of connectors to adjacent levels. The aim of this study was to determine the stiffness of side-to-side and end-to-end connectors for comparison with the use of continuous rods.
Methods:
Ten human lumbar spine specimens (L1-S1) were tested about the three axes under pure moment loading of ± 7.5 Nm. Nine conditions were used to investigate the functions of the extensions for different instrumentation lengths (L3-S1 and L2-S1) and different connector levels (L3/4 and L2/3). The intersegmental range of motion (iROM) and intersegmental neutral zone as well as total range of motion (tROM) and total neutral zone (tNZ) were analyzed.
Results:
The application of the spinal system significantly decreased the tROMs (- 44 to - 83%) and iROMs in levels L2/3 (- 56 to - 94%) and L3/4 (- 68 to - 99%) in all the tested directions, and the tNZ under flexion/extension (- 63 to - 71%) and axial rotation (- 34 to - 72%). These decreases were independent of the employed configuration (p < 0.05). The only significant changes in the iROM were observed under lateral bending between the continuous rod and the side-to-side connector at level L3/4 (p = 0.006).
Conclusion:
From a biomechanical viewpoint, the tested connectors are comparable to continuous rods in terms of ROM and NZ. These slides can be retrieved under Electronic Supplementary Material.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Instrument Transformers
Uncertainty in Measurement: Reading Instruments
Capillary Electrophoresis: Instrumentation
Fluorescence and Phosphorescence: Instrumentation