Subclinical cardiac perforation caused by a Micra™ leadless pacemaker
Ikuko Togashi1, Toshiaki Sato2, Kyoko Hoshida1
1Cardiology Kyorin University School of Medicine Mitaka Japan.
Abstract:
A subclinical cardiac perforation by a device cup of the Micra™ transcatheter pacing system was suspected in a 78-year-old woman. During the procedure, the device cup was placed on the septum. The contrast media was injected before device deployment and remained outside of the myocardium. Later, a cardiac computed tomography scan visualized a protruded diverticular structure on the right ventricle. The contrast material remained in a pouch within the pericardium. To ensure the device is oriented away from the border between the right ventricular septum and the free wall, right anterior oblique view should be carefully reviewed before deployment.
More Related Videos
Related Concept Videos
Esophageal Perforation-I: Introduction
The location of esophageal perforation can vary, occurring anywhere along the esophagus....
Esophageal Perforation-II: Clinical Manifestations and Management
Clinical Manifestations:
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
The Cardiac Cycle
The Process
Electrical signals—sent from the sinoatrial (SA) node in the right atrial wall to the atrioventricular (AV) node between the right atrium and right ventricle—cause both atria to simultaneously contract. When the signal reaches the AV node, it pauses for approximately a tenth of a second, allowing the atria to contract and...
Cardiac Cycle
During the cardiac cycle, blood flow through the heart is regulated entirely by changing pressure gradients. This sequence of events begins with the heart in a state of total relaxation, known as mid-to-late diastole, during which blood passively flows from...