Unique static discharge events on two CentriMag® (2nd Gen) ECMO patients causing console failure
Robert K Wise1, Sherry M Waldhausen1, Van T Nguyen2
11 Penn State Hershey Medical Center, Penn State Health, Hershey, PA, USA.
Two CentriMag console failures during patient transport were linked to a non-approved mounting rod causing static discharge. Removing the rod resolved the issue, preventing patient complications and further pump failures.
Area of Science:
- Cardiovascular Medicine
- Medical Device Engineering
Background:
- Extracorporeal membrane oxygenation (ECMO) use for cardiovascular support is increasing globally.
- ECMO therapy is associated with potential serious complications, including device malfunctions.
More Related Videos
04:20Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
04:05Author Spotlight: Unveiling Prognostic Indicators in Heart Failure - The Role of Phase Angle and Bioelectrical Impedance Analysis
Published on: June 30, 2023
Related Concept Videos
Other Unique Bacteria
Discharge Summary Forms
Here's a detailed look at the key components and guidelines for preparing a discharge summary:
Second Uniqueness Theorem
In contrast, consider that the electric field is non-unique and apply Gauss's law in divergence form in the region between the conductors and the integral form to the surface...
Design Example: Vintage Mixing Console
The specifications for the pre-amplifier were clear. It needed to amplify the audio signal by a factor of 10, have an input impedance above 10...
Static Equilibrium - I
Static Equilibrium - II