Randomized study of digital single-operator cholangioscope compared to fiberoptic single-operator cholangioscope in a
Raj J Shah1, Horst Neuhaus2, Mansour Parsi3
1Gastroenterology, University of Colorado, Aurora, Colorado, United States.
Background And Study Aims:
Cholangiopancreatoscopy is utilized for diagnosis and therapy of pancreaticobiliary disorders. a fully-disposable, digital, single-operator cholangioscope (DSOC) was developed with high image resolution and wide field-of-view. This bench study compared the new DSOC to the previous semi-disposable, fiber-optic cholangioscope (FSOC) prior to the clinical availability of the DSOC system.
Methods:
Five experts performed one practice run followed by randomized runs comparing DSOC to FSOC in a biliary tract model consisting of three fixed left-intrahepatic tracts (LIHD), and variable common bile duct (CBD) and right-intrahepatic tracts (RIHD) with seven total lesions in multiple configurations. Timed runs aimed to visualize and target each lesion using miniature biopsy forceps. Definitions: visual success, visualizing targets; targeting success, touching target with forceps; complete run, touching seven targets within 20 minutes. Image quality, ease-of-use, and time to completion were recorded.
Results:
Thirty-seven evaluable runs (20 DSOC, 17 FSOC) were completed. DSOC was superior to FSOC in Visual (99 % vs. 67 %, P < 0.001) and targeting success (6.6 vs. 4.5, P = 0.009), proportion of complete runs (13 /20 vs. 0 /17, P < 0.001) and time of run (10.1 min vs. 15.4 min, P < 0.001). For fixed LIHD, DSOC achieved higher targeting success compared to FSOC (2.6 vs. 1.1, P < 0.001) with no difference in RIHD and CBD targets (4.0 vs. 3.4, P = 0.39). Investigators reported superior image quality and ease-of-use with DSOC.
Conclusions:
In this model, DSOC performed superiorly to FSOC in image quality, visualization, and maneuverability. The model could potentially be utilized for training endoscopists less experienced with cholangiopancreatoscopy.
Related Concept Videos
Operational Amplifiers
Real Number Operations
Vector Operations
A vector multiplied by a scalar value is called scalar multiplication. The result obtained is a new vector with a different magnitude. If the scalar is positive, the direction of the vector remains the same, but if it is negative, the direction of the vector is reversed. For example, the product of the mass and velocity yields the momentum.
Operant Conditioning
Reinforcement in operant conditioning can be positive or negative, both of which serve to increase the likelihood of a behavior. Positive...
Second Derivatives and Laplace Operator
Consider a scalar function. The curl of its...
Modes of Operations of BJT
Active Mode: The most common mode for amplification, the active mode features a forward-biased emitter-base junction and a reverse-biased base-collector junction. This setup enables electrons to be injected from the emitter to the base while blocking the majority carriers at the collector. The...