A precise, sensitive and stable LC-MSMS method for detection of picomolar levels of serum aldosterone
Margrete Lie1, Ketil Thorstensen1
1a Department of Medical Biochemistry , St Olavs Hospital, Trondheim University Hospital , Trondheim , Norway.
Background:
Accurate quantification of aldosterone is essential in the diagnosis of primary aldosteronism. Liquid chromatography-tandem mass spectrometry (LC-MSMS) analysis is increasingly being used to improve analytical sensitivity and specificity, since this technology reduces most of the interferences observed with immunological methods.
Methods:
Serum samples with d7-aldosterone as internal standard were extracted with methyl tert-butyl ether, using liquid-liquid extraction (LLE). Chromatographic separation was performed on a C18 reverse phase column with a methanol-water gradient containing ammonium fluoride. Aldosterone detection was performed on an Agilent 6490 triple quadrupole using electro spray ionisation in positive mode.
Results:
Multiple reaction monitoring transitions were m/z 361.2-315.1 for aldosterone, and 368.5-323.3 for d7-aldosterone. Chromatographic retention time was 2.7 min. The method's total CVs at aldosterone concentrations of 45.4 and 1080 pmol/L were 7.0% and 4.8%, respectively. The intra-assay CVs at concentrations of 60.0 and 637 pmol/L were 4.0% and 2.6%, respectively. The method's LOQ and LOD were 10 and 5 pmol/L, respectively, demonstrating an excellent analytical sensitivity. The upper limit of quantification was set to 5000 pmol/L, corresponding to the highest calibrator concentration. The long-term stability of the method was evident from repeated measurements of external control pools from UKNEQAS over a period of about 3 years, showing CVs between 2.0 and 7.0%.
Conclusions:
We have described a precise, sensitive and stable LC-MSMS method for the measurement of serum aldosterone. In addition, due to the use of LLE and a short LC-column, the method is simple to perform, with a short chromatographic run time.
Related Concept Videos
Introduction and Methods of Leveling
Uncertainty in Measurement: Accuracy and Precision
Leveling Effect
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...