Liver MR Elastography at 3 T: Agreement Across Pulse Sequences and Effect of Liver R2* on Image Quality
Ely R Felker1, Kang-Sun Choi2, Kyung Sung1
11 Department of Radiology, Ronald Reagan-UCLA Medical Center, 757 Westwood Blvd, Los Angeles, CA 90095.
Objective:
The objectives of our study were to compare MR elastography (MRE) based on gradient-recalled echo (GRE) imaging with spin-echo echo-planar imaging (SEEPI) and rapid fractional (RF)-GRE MRE sequences at 3 T in terms of liver stiffness (LS) and image quality and to evaluate the effect of liver R2* on image quality.
Materials And Methods:
Eighty-one patients underwent 3-T liver MRE with GRE, SE-EPI, and RF-GRE sequences performed in variable order in this study. LS and ROI areas on the LS 95% confidence maps were compared among the three sequences. The relationship between liver R2* and ROI area was investigated.
Results:
There was no significant difference in mean LS among the three sequences (p = 0.49). Mean ROI area was significantly larger for RF-GRE (18,213 ± 9292 [SD] mm2) than for GRE (13,196 ± 8149 mm2) and SE-EPI (12,896 ± 8656 mm2) (p < 0.0001). Liver R2* was significantly higher among patients with one or more failed sequences (mean ± SD, 116 ± 76 s-1) than for patients with no failed sequences (59 ± 26 s-1) (p = 0.001). Technical failure rates were 10% (8/81), 4% (3/81), and 2% (2/81) for GRE, SE-EPI, and RF-GRE, respectively. Among patients with iron overload (R2* ≥ 100 s-1), there was a trend toward larger ROI area for SE-EPI (p = 0.09).
Conclusion:
SE-EPI-and RF-GRE-based MRE sequences provide equivalent measures of LS compared with GRE-based MRE, and both have lower technical failure rates. The RF-GRE sequence yielded the largest measurable area of LS. Among patients with iron overload, there was a trend toward larger measurable area of LS for the SE-EPI sequence.
More Related Videos
09:27Analysis of Liver Microenvironment During Early Progression of Non-Alcoholic Fatty Liver Disease-Associated Hepatocellular Carcinoma in Zebrafish
Published on: April 1, 2021
05:25Author Spotlight: Advancements and Challenges in Machine Perfusion for Liver Transplantation
Published on: June 14, 2024
Related Concept Videos
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Liver Histology
Hepatocytes perform a variety of essential functions. They secrete...
Liver Physiology
Metabolic Regulation:
The liver is the central organ involved in regulating blood composition. It stabilizes blood glucose levels, maintaining them within the range of 70–110 mg/dL. When these levels drop, the liver breaks down glycogen reserves and releases glucose into the bloodstream. It can...
Liver Regeneration
Cells of Liver
The liver comprises four major types of cells— hepatocytes, stellate, Kupffer, and sinusoidal endothelial cells. The hepatocytes are...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Gross Anatomy of the Liver
Located under the diaphragm, the liver is almost entirely ensconced within the rib cage, providing it with substantial protection. Except for the superior most bare area, the liver's surface is...