Related Experiment Video
Updated: Feb 7, 2026

Toxicity Screens in Human Retinal Organoids for Pharmaceutical Discovery
Published on: March 4, 2021
Retinal Toxicity of Medical Devices Used during Vitreoretinal Surgery: A Critical Overview
Kai Januschowski1,2, Cristina Irigoyen3, J Carlos Pastor4,5,6,7
1Eye Clinic Sulzbach, Knappschaftshopsital Sulzbach/Saar, Sulzbach-Saar, Germanykai.januschowski@kksaar.de.
Abstract:
Retinal toxicity/biocompatibility of medical devices in direct contact with the retina is an important subject for clinicians and scientists. As these effects are not very frequent, there is also a relative lack of information for many clinicians. The past has taught us multiple times that there is a significant safety problem associated with severe loss of vision in affected patients. In this review, we want to classify medical products that are used in the back of the eye, describe recent examples of toxicity, critically reflect on the regulations that exist and suggest improvements that can be done to ensure patient safety without hindering innovation.
Methods:
Critical review of the recent papers and personal experience of the authors in this issue. Medical devices used in the back of the eye and recent examples of toxicity are described, regulations that exist are critically reflected and improvements suggested that can ensure patient safety without hindering innovation.
Results:
There is clear evidence of toxicity after intraocular surgery in any category. Some cytotoxic indirect methods have failed in detecting this toxicity. Some ISO rules do not seem appropriate. Postmarketing safety is missing. There is little data on this issue.
Conclusions:
The absence of a clear regulation of the production, purification and evaluation of the toxic effects of the medical devices supposes the possibility that products are not sufficiently safe to obtain the CE mark.
Related Concept Videos
Toxic Reactions: Overview
Toxicity falls into two primary categories: local and systemic.
Local toxicity appears at the exposure site, such as protein denaturation caused by caustic substances.
In contrast, systemic toxicity requires the toxic agent's absorption and distribution,...
Critical Region, Critical Values and Significance Level
In hypothesis testing, a sample statistic is converted to a test statistic using z, t, or chi-square distribution. A critical region is an area under the curve in probability distributions demarcated by the critical value. When the test statistic falls in this region, it suggests that the null hypothesis must be rejected. As this region contains all those values of the...
Critical Values
Inhaled Medications
Critical Thinking I
Critical Thinking II