A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Evaluation of a novel automatic screening tool for determining eligibility for a subcutaneous implantable
Rafi Sakhi1, Sing-Chien Yap1, Michelle Michels1
1Department of Cardiology, Erasmus Medical Center, Rotterdam, the Netherlands.
Background:
The manufacturer has developed a new ECG screening tool to determine eligibility for the subcutaneous ICD (S-ICD), the "automatic screening tool" (AST), which may render manual ECG-screening unnecessary. The aim of the study was to determine the eligibility for the S-ICD using two methods (manual ECG-screening versus AST) in different patient categories including patients with cardiomyopathy, congenital heart disease and inherited primary arrhythmia syndrome.
Methods:
We prospectively evaluated the ECG suitability for an S-ICD in consecutive patients at our outpatient clinic between February and June 2017. The primary endpoint of the study was ECG eligibility defined as at least 1 successful vector in both supine and sitting postures.
Results:
A total of 254 patients (167 men; mean age 45 ± 16 years) were screened using both methods. Overall, there was a high ECG eligibility using either method (93% versus 92%, P = 0.45). Overall agreement between both methods was 94%. Patients with hypertrophic cardiomyopathy (HCM) more often had a failed screening test using either test in comparison to the patients without HCM (manual: odds ratio [OR] 3.3, 95% confidence interval [CI] 1.2-9.3, P = 0.02; AST: OR 3.0, 95% CI 1.2-7.6, P = 0.02).
Conclusion:
AST showed a high agreement with manual ECG-screening for S-ICD. Overall there was a high ECG eligibility for S-ICD, although patients with HCM had a lower passing rate irrespective of the screening method.
More Related Videos
08:38Renal Capsule Xenografting and Subcutaneous Pellet Implantation for the Evaluation of Prostate Carcinogenesis and Benign Prostatic Hyperplasia
Published on: August 28, 2013
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
Automatic Processing and Automatic Social Behavior
Determining Order of Reaction
Determination
Determining the pH of Salt Solutions
Genetic Screens
Forward genetic screens
Forward or “classical” genetic screens involve creating random mutations in an organism’s DNA using radiation, mutagens, or insertion of additional bases, which...
Rate-Determining Steps
In a multistep reaction mechanism, one of the elementary steps progresses significantly slower than the others. This slowest step is called the rate-limiting step (or rate-determining step). A reaction cannot proceed faster than its slowest step, and hence, the rate-determining step limits the overall reaction rate.
The concept of rate-determining step can be understood from the analogy of a 4-lane freeway with a short-stretch of traffic-bottleneck caused due to...