Dual layer computed tomography: Reduction of metal artefacts from posterior spinal fusion using virtual monoenergetic
Julia Dangelmaier1, Benedikt J Schwaiger1, Alexandra S Gersing1
1Department of Diagnostic and Interventional Radiology, Technische Universität München, Ismaningerstr. 22, 81675, Munich, Germany.
Introduction:
To evaluate the clinical potential of dual layer computed tomography (DLCT) for posterior fusions of the thoracic and lumbar spine and determine the optimal keV-settings for an improved overall image quality and effective reduction of metal artefacts affecting the implant inheriting vertebral body, the spinal canal, the paravertebral muscle and aorta.
Methods And Materials:
Twenty patients with posterior thoracic and lumbar spinal fusion, who underwent a 120kVp- DLCT scan were included in this study. Two independent readers evaluated axial 0.9 mm slides with soft tissue and bone window settings. Image quality of the conventional scan was compared to virtual monoenergetic images (VMI) at 40, 60, 80, 100,120, 140, 160, 180 and 200 keV. Diagnostic image quality was assessed on a four point Likert-scale overall, as well as specifically for the implant inheriting bone, paravertebral muscle, spinal canal or aorta. The Hounsfield Units (HU) of the area with the most pronounced streak artefact as well as HU of a reference area containing fat and muscle were documented for each keV-setting and compared to the conventional image. SNR and CNR were calculated for each of the four anatomic areas. Statistical analysis was conducted for the total collective and separately for the thoracic and lumbar spine level.
Results:
Starting from 80 keV qualitative analysis revealed significant improvement of overall image quality and benefit for each tissue separately compared to the conventional images (CI) (p-values in the range from <0.001 to 0.005). 180 keV was considered the optimal monoenergetic setting regarding the overall image quality. For the assessment of the implant inheriting bone, the spinal canal, paravertebral muscle and aorta 200, 180, 160 and 180 keV were rated to be the most sufficient. Our results reveal high inter-reader agreement for qualitative evaluations (intra-class correlation coefficients >0.927; p < 0.05). HU values within the most pronounced streak artefact increased significantly with higher keV (p < 0.001), while there was no significant alteration of HU within the reference area. A decrease in SNR and CNR for higher VMI was revealed by our results.
Conclusion:
VMIs of higher energies provide significant reduction of metallic artefacts from posterior spinal fusions. Dedicated keV settings to evaluate either the implant inheriting bone, the spinal canal,adjacent muscle or aorta - structures, which are frequently of particular interest after posterior spinal fusion - are recommended. In addition, an optimal keV for an improved overall image quality is proposed.
More Related Videos
03:58Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
09:20Novel Photoacoustic Microscopy and Optical Coherence Tomography Dual-modality Chorioretinal Imaging in Living Rabbit Eyes
Published on: February 8, 2018
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Nuclear Fusion
A helium nucleus has a mass that is 0.7% less than that of four hydrogen nuclei; this lost mass is converted into energy during the fusion. This reaction produces about...
Oxidation-Reduction Reactions
Metallic Solids
All metallic solids exhibit high thermal and electrical conductivity, metallic luster, and malleability....
Metal-Ligand Bonds
In these complexes, transition metals form coordinate covalent bonds, a kind of Lewis acid-base interaction in which both of the electrons in the bond are contributed by a donor (Lewis base) to an electron acceptor (Lewis acid). The Lewis acid in...