Annalisa Patriarca1, Charles Fouillade2, Michel Auger1
1Institut Curie, PSL Research University, Radiation Oncology Department, Paris, France.
Purpose:
Recent in vivo investigations have shown that short pulses of electrons at very high dose rates (FLASH) are less harmful to healthy tissues but just as efficient as conventional dose-rate radiation to inhibit tumor growth. In view of the potential clinical value of FLASH and the availability of modern proton therapy infrastructures to achieve this goal, we herein describe a series of technological developments required to investigate the biology of FLASH irradiation using a commercially available clinical proton therapy system.
Methods And Materials:
Numerical simulations and experimental dosimetric characterization of a modified clinical proton beamline, upstream from the isocenter, were performed with a Monte Carlo toolkit and different detectors. A single scattering system was optimized with a ridge filter and a high current monitoring system. In addition, a submillimetric set-up protocol based on image guidance using a digital camera and an animal positioning system was also developed.
Results:
The dosimetric properties of the resulting beam and monitoring system were characterized; linearity with dose rate and homogeneity for a 12 × 12 mm2 field size were assessed. Dose rates exceeding 40 Gy/s at energies between 138 and 198 MeV were obtained, enabling uniform irradiation for radiobiology investigations of small animals in a modified clinical proton beam line.
Conclusions:
This approach will enable us to conduct FLASH proton therapy experiments on small animals, specifically for mouse lung irradiation. Dose rates exceeding 40 Gy/s were achieved, which was not possible with the conventional clinical mode of the existing beamline.
Related Concept Videos
Cavity Drainage and Flashings in Masonry walls
Weep holes, strategically placed at the base of the cavity, are critical for draining accumulated water. These openings are created by leaving head...
Proton (¹H) NMR: Chemical Shift
Absorption signals of all the protium nuclei...
Setting Time of Cement
¹H NMR of Labile Protons: Temporal Resolution
The –OH proton in alcohols typically appears in the range of δ 2 to 5 ppm but can vary depending on the specific...
¹H NMR of Labile Protons: Deuterium (²H) Substitution
Tonicity in Animals