[Dose distribution of electrons obtained from linear accelerators equipped with a beam scanning system]
Abstract:
Homogeneous distribution of electrons used for therapeutic purposes and obtained from accelerators, is achieved by means of Potter-Bucky diaphragms or by repeated, staggered, sawtooth-shaped sweeping movements of the electron beam (scanning) over the radiation field. The repetition of the scanning process (number of scans) can result in long measurement times for achieving a sufficiently homogeneous, dosimetrically adequate distribution of the electrons. This "time problem" makes it imperative to achieve good homogeneity while keeping the number of scans as low as possible. To solve the problem, the scanning movement of the electron beam is simulated by a computer programme and the interdependence of the homogeneity of the irradiation field and number of scans is investigated. Since changing the ratio of the two deflection rates exercises a significant influence, it is mandatory in dosimetry to pay close attention to a strict observance of the deflection rates.