Retinal Vascular Reactivity as Assessed by Optical Coherence Tomography Angiography
Published on: March 26, 2020
Repeatability and Reproducibility of Quantitative Assessment of the Retinal Microvasculature Using Optical Coherence
Qi Zhao1, Wen Li Yang1, Xiao Na Wang2
1Department of Ophthalmology, Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University; Beijing Key Laboratory of Ophthalmology and Visual Sciences, Beijing 100730, China.
Objective:
The aim of this study was to determine the repeatability and reproducibility of optical coherence tomography angiography (OCTA) based on optical microangiography (OMAG) measurements of macular vessels in normal eyes.
Methods:
In this prospective cohort study, 40 eyes of 40 healthy volunteers underwent repeated OCTA (Cirrus HD-OCT 5000 angiography system, Carl Zeiss Meditec, Inc.) scans on two separate visit days. On each visit day, the eyes were scanned three times. The following parameters were used to quantitatively describe the OCTA images of the superficial vascular network: vessel area density (VAD), vessel skeleton density (VSD), vessel diameter index (VDI), vessel perimeter index (VPI), vessel complexity index (VCI), flux, and foveal avascular zone (FAZ). Coefficient of variation (CV) and intraclass correlation coefficient (ICC) were calculated for evaluating intravisit and intervisit repeatability, as well as interobserver reproducibility.
Results:
The measurements showed high repeatability [CVs ⪕ 4.2% (intravisit) and ⪕ 4.6% (intervisit)] and interobserver reproducibility (ICCs ⪖ 0.923) for all parameters.
Conclusion:
This study demonstrated good repeatability and reproducibility of OCTA based on OMAG for the measurement of superficial vessel parameters in normal eyes.
Related Concept Videos
Properties of Enantiomers and Optical Activity
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Quantitative Analysis
In quantitative analysis, two key measurements are made: the sample quantity and a property proportional to the amount of the analyte (the substance being analyzed). This forms the basis of the...
Imaging Studies III: Computed Tomography