Pressure-Volume Loop Analysis of Multipoint Pacing With a Quadripolar Left Ventricular Lead in Cardiac
Wouter M van Everdingen1, Alwin Zweerink2, Odette A E Salden1
1Department of Cardiology, University Medical Center Utrecht, Utrecht, the Netherlands.
Objectives:
This study aimed to compare multipoint pacing (MPP) to optimal biventricular pacing with a quadripolar left ventricular (LV) lead and find factors associated with hemodynamic response to MPP.
Background:
MPP with a quadripolar LV lead may increase response to cardiac resynchronization therapy.
Methods:
Heart failure patients with a left bundle branch block underwent cardiac resynchronization therapy implantation. Q to LV sensing interval divided by the intrinsic QRS duration was measured. Invasive pressure-volume loops were assessed during 4 biventricular pacing settings and 3 MPP settings, using 4 atrioventricular delays. Hemodynamic response was defined as change in stroke work (Δ%SW) compared with baseline measurements during intrinsic conduction. Δ%SW of MPP was compared with conventional biventricular pacing using the distal electrode and the electrode with highest Δ%SW (BIV-OPT).
Results:
Forty-three patients were analyzed (age 66 ± 10 years, 63% men, 30% ischemic cardiomyopathy, LV ejection fraction 29 ± 8%, and QRS duration 175 ± 13 ms). Q to local LV sensing interval corrected for QRS duration was 84 ± 8%, and variation between LV electrodes was 9 ± 5%. Compared with conventional biventricular pacing using the distal electrode, MPP showed a significant higher increase of SW (Δ%SW +15 ± 35%; p < 0.05) with a large interindividual variation. There was no significant difference in Δ%SW with MPP compared with BIV-OPT (-5 ± 24%; p = 0.19). Male sex and low LV ejection fraction were associated with increase in Δ%SW due to MPP versus BIV-OPT in multivariate analysis, while ischemic cardiomyopathy was only associated in univariate analysis.
Conclusions:
Optimization of the pacing site of a quadripolar LV lead is more important than to program MPP. However, specific subgroups (i.e., especially men) may benefit substantially from MPP.
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Le Chatelier's Principle: Changing Volume (Pressure)
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Pressure and Volume in an Adiabatic Process
Cardiac Catheterization III: Left Heart Catheterization
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer