Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Ambient dose during intraoral radiography with current techniques: Part 1 conversion factor for scattered radiation
Reinier C Hoogeveen1, Tyrell I de Randamie1, Gianni M Soemodihardjo1
11 Department of Maxillofacial Radiology of the Academic Centre For Dentistry Amsterdam (ACTA), University of Amsterdam and VU University Amsterdam , Amsterdam , The Netherlands.
A new conversion factor for scattered radiation in intraoral radiography was established. This factor is three times lower than previously reported values, improving radiation safety assessments.
Area of Science:
- Medical Physics
- Radiological Protection
- Dental Imaging
Background:
- Intraoral radiography utilizes rectangular collimators to minimize patient radiation exposure.
- Assessing scattered radiation is crucial for accurate dose monitoring and radiation safety protocols.
- Existing conversion factors may not fully represent scattered radiation in specific intraoral radiography scenarios.
Purpose of the Study:
- To establish a conversion factor for scattered radiation during intraoral radiography using a rectangular collimator.
- To quantify scattered radiation dose in areas outside the primary X-ray beam.
- To provide a more accurate factor for radiation safety assessments in dental settings.
Main Methods:
- Utilized a Rando phantom exposed to various geometries with a Planmeca Intra X-ray unit.
- Measured scattered radiation dose at multiple horizontal and vertical locations.
- Corrected measurements for background and leakage radiation, and weighted them by exposure workload.
Main Results:
- Scattered ambient radiation dose showed even distribution in the horizontal plane.
- Dose in the vertical plane was 20-65% lower compared to the horizontal plane.
- A representative scattered radiation pattern was constructed based on weighted exposures.
Conclusions:
- Established a conversion factor of 0.05 µSv per mAs at 1 meter for scattered ambient radiation.
- This new factor is approximately three times lower than previously reported values.
- The findings contribute to improved radiation safety and dose assessment in intraoral radiography.
More Related Videos
Related Concept Videos
Biological Effects of Radiation
Gene Conversion
Gene Conversion
Curvilinear Motion: Rectangular Components
As the car advances, its position evolves over time. Quantifying the car's velocity involves computing the...
Conversion of Units
The first step in the unit conversion is to list the given units and the units required...
Rectangular and Triangular Pulse Function
For example, consider a rectangular pulse with a 5V amplitude, a 3-second duration, and centered at t=2 seconds. This pulse can be expressed using the rectangular function, written as,