Systolic time intervals assessed from analysis of the carotid pressure waveform
Paolo Salvi1, Andrea Grillo1,2, Isabella Tan3
1Istituto Auxologico Italiano, IRCCS, Department of Cardiovascular, Neural and Metabolic Sciences, Milan, Italy.
Objective:
The timing of mechanical cardiac events is usually evaluated by conventional echocardiography as an index of cardiac systolic function and predictor of cardiovascular outcomes. We aimed to measure the systolic time intervals, namely the isovolumetric contraction time (ICT) and pre-ejection period (PEP), by arterial tonometry.
Approach:
Sixty-two healthy volunteers (age 47 ± 17 years) and 42 patients with heart failure and reduced ejection fraction were enrolled (age 66 ± 14 years). Pulse waves were recorded at the carotid artery by arterial tonometry together with simultaneous aortic transvalvular flow by Doppler-echocardiography, synchronized by electrocardiographic gating. The ICT was determined from the time delay between the electrical R wave and the carotid pressure waveform, after adjustment for the pulse transit time from the aortic valve to the carotid artery site, estimated by an algorithm based on the carotid-femoral pulse wave velocity. The PEP was evaluated by adding the electrical QR duration to the ICT.
Main Results:
The ICT derived from carotid pulse wave analysis was closely related to that measured by echocardiography (r = 0.90, p < 0.0001), with homogeneous distribution in Bland-Altman analysis (mean difference and 95% confidence interval = 0.2 from -14.2 to 14.5 ms). ICT and PEP were higher in cardiac patients than in healthy volunteers (p < 0.0001). The ratio between PEP and left ventricular ejection time was related to the ejection fraction measured with echocardiography (r = 0.555, p < 0.0001).
Significance:
The timing of electro-mechanical cardiac events can be reliably obtained from the carotid pulse waveform and carotid-femoral PWV, evaluated using arterial tonometry. Systolic time intervals assessed with this approach showed good agreement with measurements performed with conventional echocardiography and may represent a promising additional application of arterial tonometry.
Related Concept Videos
Prediction Intervals
However, the point estimate is most likely not the exact value of the population parameter, but close to it. After calculating point estimates, we construct interval estimates, called confidence intervals or prediction intervals. This prediction interval comprises a range of values unlike the point estimate and is a better predictor of the observed sample value, y.
Effective Value of a Periodic Waveform
The effective value of a periodic current represents the direct current (DC) that conveys the same average power to a resistor as the periodic current itself. This concept is crucial when assessing AC circuits. To determine the...
Assessing Blood pressure in the Leg
Preparation:
Confidence Intervals
A...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Improper Integrals: Infinite Intervals