[Exposure Dose Reduction for Radiologists with Combination of Angular Beam Modulation and Radiation Protection Drape
Yuko Seki1, Yasuhiro Fukushima1,2, Masashi Ando1
1Department of Radiology, Gunma University Hospital.
Purpose:
To assess the dose reduction of radiologists by using angular beam modulation (ABM) and radiation protection drape during computed tomography (CT) fluoroscopy.
Materials And Methods:
The phantom was set on the lower that is 15 cm from the isocenter position. We measured the radiation exposure around the phantom with radiophotoluminescence glass dosimeters. The space radiation dose rate was measured with an ionization chamber dosimeter in the CT room.
Results:
The dose rate of finger radiation exposure was 67% at assumed assist tool position with ABM. And the dose rate of finger radiation exposure with the combination of ABM and radiation protection drape was 33%. The space dose rate of exposure with the combination of ABM and radiation protection drape was 49% at 150 cm.
Conclusion:
The combination of ABM and radiation protection drape can reduce finger radiation exposure at assumed assist tool position. The space dose rate of the standing position of radiologists can get a clear dose of radiation reduction by the combination of both.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Related Concept Videos
Biological Effects of Radiation
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Bioavailability Study Design: Single Versus Multiple Dose Studies
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...