[Enhancement of Pesticide Peak Response in GC-MS in the Presence of Multiple Co-Existing Reference Pesticides]
Masato Yoshimitsu1, Kazuhiko Akutsu1, Yoko Kitagawa1
1Osaka Institute of Public Health.
Abstract:
We carried out a collaborative study in six laboratories to confirm the universality of the enhancing effect of co-existing reference pesticides on the GC-MS peak response to a target pesticide (malathion, procymidone, or flucythrinate). First, we confirmed the response enhancement of the target pesticides with increasing numbers of co-existing reference pesticides in solution. Then, using diluted green soybean matrix, we analyzed the target pesticides with two types of matrix-matched calibration, containing the target pesticides or 166 other pesticides. In both cases, the response-enhancing effect of co-existing pesticides was confirmed in all laboratories. The enhancement was reduced by addition of green soybean matrix to the sample and calibration solutions. Our results show that it is necessary to estimate the peak response-enhancing effect of co-existing pesticides in the calibration solution to obtain accurate results with GC-MS determination. The enhancing effect could be reduced by addition of food matrix to the sample and calibration solutions.
More Related Videos
08:22Author Spotlight: A New and Efficient Method for Comprehensive Metabolite Cytotoxicity Assessment of Triazole Pesticides in Plants
Published on: December 22, 2023
04:17Analysis of Organochlorine Pesticides in a Soil Sample by a Modified QuEChERS Approach Using Ammonium Formate
Published on: January 20, 2023
Related Concept Videos
Inertial Frames of Reference
Non-inertial Frames of Reference
IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations
Self-Evaluation: Self-Enhancement and Self-Verification
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...