Monitoring of tidal ventilation by electrical impedance tomography in anaesthetised horses
M Mosing1, A D Waldmann2,3, A Raisis1
1College of Veterinary Medicine, Murdoch University, Perth, Australia.
Background:
Electrical impedance tomography (EIT) is a method to measure regional impedance changes within the thorax. The total tidal impedance variation has been used to measure changes in tidal volumes in pigs, dogs and men.
Objectives:
To assess the ability of EIT to quantify changes in tidal volume in anaesthetised mechanically ventilated horses.
Study Design:
In vivo experimental study.
Methods:
Six horses (mean ± s.d.: age 11.5 ± 7.5 years and body weight 491 ± 40 kg) were anaesthetised using isoflurane in oxygen. The lungs were mechanically ventilated using a volume-controlled mode. With an end-tidal carbon dioxide tension in the physiological range, and a set tidal volume (VTvent ) of 11-16 mL/kg (baseline volume), EIT data and VT measured by conventional spirometry were collected over 1 min. Thereafter, VTvent was changed in 1 L steps until reaching 10 L. After, VTvent was reduced to 1 L below the baseline volume and then further reduced in 1 L steps until 4 L. On each VT step data were recorded for 1 min after allowing 1 min of stabilisation. Impedance changes within the predefined two lung regions of interest (EITROI ) and the whole image (EITthorax ) were calculated. Linear regression analysis was used to assess the relationship between spirometry data and EITROI and EITthorax for individual horses and pooled data.
Results:
Both EITROI and EITthorax significantly predicted spirometry data for individual horses with R2 ranging from 0.937 to 0.999 and from 0.954 to 0.997 respectively. This was similar for pooled data from all six horses with EITROI (R2 = 0.799; P<0.001) and EITthorax (R2 = 0.841; P<0.001).
Main Limitations:
The method was only tested in healthy mechanically ventilated horses.
Conclusions:
The EIT can be used to quantify changes in tidal volume.
More Related Videos
08:06Author Spotlight: Optimizing Porous Substrate Electroporation Through Micro and Nanochannels for Enhanced Monitoring and Intermediate Stage Characterization
Published on: September 27, 2024
05:44Performing In Vivo and Ex Vivo Electrical Impedance Myography in Rodents
Published on: June 8, 2022
Related Concept Videos
Tidal Forces
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Impedance Combination
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...