Repeatability and reproducibility of retinal and choroidal thickness measurements in Diabetic Macular Edema using
Anna Sala-Puigdollers1,2,3, Marc Figueras-Roca1,2, Mireia Hereu1
1Institut Clínic d'Oftalmologia (ICOF), Hospital Clínic, Barcelona, Spain.
Purpose:
To evaluate the repeatability and reproducibility of retinal and choroidal thickness measured with Swept source Optical Coherence Tomography (SS-OCT) in eyes with Diabetic Macular Edema (DME).
Methods:
42 DME eyes were imaged using SS-OCT standard Macular scanning protocols. Retinal and choroidal thickness were measured in the Total macular circle (TMC) and foveal central subfield (FCS) using device-integrated specific software. The coefficient of repeatability (CR) and intraclass correlation coefficient (ICC) were determined as a measure of repeatability and relative reliability within graders. Reproducibility was assessed using Bland-Altman plots and 95% limits of agreement (LoA) were determined as a measure of interobserver variability.
Results:
Intragrader CR of retinal and choroidal thickness were 8.37 and 12.20 microns for TMC and 22.24 and 32.40 microns for FCS, and intergrader 95% LoA were 7.37-8.69 and -27.2-27.71 microns for TMC and -34.21-41.93 and -30.46-24.84 for FCS, respectively. Retinal and choroidal thickness showed very good intraobserver reliability for both TMC and FCS (ICC 0.99, LoA 0.98-0.99 in all cases). Intraobserver and interobserver variability for retinal and choroidal thickness was not significantly different for TMC (p = 0.98 and p = 0.90, p = 0.98 and p = 0.91) or FCS (p = 0.97 and p = 0.85, p = 0.78 and p = 0.73), respectively.
Conclusions:
Retinal and choroidal thickness in DME eyes can be quantified with good reliability, repeatability and reproducibility using new OCT devices that incorporate swept source technology. The technical advantages of this technology may provide new insights in the understanding of the choroidal changes related with DME.
More Related Videos
Related Concept Videos
Pathophysiology of Diabetes
Type 1 diabetes is characterized by autoimmune-mediated destruction of pancreatic β cells, with environmental factors potentially triggering this process in genetically susceptible individuals. Despite many not having a family history, certain genes increase susceptibility,...
Bioequivalence Experimental Study Designs: Repeated Measures, Cross-Over, Carry-Over, and Latin Square Designs
Sinusoidal Sources
In homes, the power supplies use sinusoidal sources to provide electricity. These sources generate a voltage that varies sinusoidally...
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Psychoneuroimmunology: Diabetes and Cancer
AC Sources