Choosing the right through-the-scope clip: a rigorous comparison of rotatability, whip, open/close precision, and closure strength (with videos)
Related Experiment Video
Updated: Feb 7, 2026

Fabricating van der Waals Heterostructures with Precise Rotational Alignment
Published on: July 5, 2019
Thomas J Wang1, Hiroyuki Aihara2, Andrew C Thompson3
1Massachusetts General Hospital, Department of Medicine; Boston, Massachusetts, USA; Brigham and Women's Hospital, Division of Gastroenterology, Hepatology, and Endoscopy, Boston, Massachusetts, USA.
Background And Aims:
Many new through-the-scope clips are available, and physicians often select clips based on physical characteristics and/or cost. However, functional profiles may be equally important and have not been methodically assessed. We evaluated 5 commercially available clips: Resolution 360, Instinct, Quick Clip Pro, Dura Clip, and SureClip.
Methods:
We rigorously compared clips on multiple characteristics, including rotatability, overshoot, open/close precision, and tensile/closure strength. Clips were tested in 4 different endoscope configurations: (1) straight, (2) duodenal sweep, (3) full retroflexion, and (4) across the duodenoscope elevator.
Results:
For rotatability, the Resolution 360 was the fastest due to its unique functionality in allowing primary MD control in rotation (P < .05). The Resolution 360, SureClip, and Dura Clip were able to rotate through the prescribed sequence across all scope configurations. For overshoot, the SureClip and Resolution 360 had the least overshoot for the straight configuration at 0%. All clips had >75% overshoot at more strained configurations. For open/close precision, the SureClip and Dura Clip showed precise opening/closing with the ability to stop at any point. The remaining clips exhibited abrupt opening with more controlled closure. For tensile strength, the Quick Clip Pro generated the highest peak force as would be required in lateral tissue manipulation (4.8 lb, P < .005). For closure strength, the Instinct overall showed the most gel compression, and along with the Resolution 360, showed 100% deployment success for all gel tissue thicknesses (up to 10 mm).
Conclusions:
Each clip has a unique physical and functional profile, which may be a factor in selection depending on the clinical circumstance.
Related Concept Videos
The Scope of Physics
Choosing Between z and t Distribution
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Acid Strength and Molecular Structure
In the absence of any leveling effect, the acid strength of binary compounds of hydrogen with nonmetals (A) increases as the H-A bond strength decreases down a group in the periodic table. For group 17, the order of increasing acidity is HF < HCl < HBr < HI. Likewise, for group 16, the order of increasing acid strength is H2O < H2S < H2Se < H2Te. Across a row in the periodic table, the acid strength of binary hydrogen compounds increases with increasing...
Uncertainty in Measurement: Accuracy and Precision
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...