Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
Effects of reducing frame rate from 7.5 to 4 frames per second on radiation exposure in transcatheter atrial septal
11Centre de Référence Malformations Cardiaques Congénitales Complexes-M3C,Hôpital Necker Enfants Malades,Assistance Publique des Hôpitaux de Paris, Paris,France.
Objectives:
The aim of this study was to evaluate the reduction of frame rate from 7.5 to 4 frames per second on radiation exposure and to provide new standards of radiation exposure.
Background:
Frame rate is a large contributor to radiation exposure. The use of 4 frames per second for closure of atrial septal defects has been reported not to affect the level of radiation exposure.
Methods:
We retrospectively reviewed radiation data from all patients referred to our catheterisation laboratory for closure of an atrial septal defect between January, 2015 and June, 2017. Fluoroscopic time, dose area product (μGy.m2), and total air kerma (mGy) were collected. These values were compared according to the frame rate used for closure of atrial septal defects.
Results:
A total of 49 atrial septal defects were closed using 7.5 frames per second and 85 using 4 frames per second. Baseline characteristics were similar in both groups. Procedural success was similar in both groups (100 versus 98.8%). Median total air kerma and dose area product were statistically lower in the 4 frames per second group (4 versus 1.3 mGy [p=0.00012]), 43.7 versus 13.1 μGy.m2 [p0.05).
Conclusion:
Reduction of frame rate allows reducing significantly the radiation exposure while maintaining excellent clinical results in transcatheter closure of atrial septal defects. We recommend implementing this little change in every laboratory in order to achieve drastic reduction of radiation exposure to the patients and laboratory personnel.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
23:33The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation
Published on: February 28, 2012
Related Concept Videos
Framing Effects
Frames
Frames are versatile and widely used in various applications such as structural supports for beams and columns, automobile chassis construction, and in the construction...
Frames: Problem Solving II
Frames: Problem Solving I
Inertial Frames of Reference
Non-inertial Frames of Reference