Reproducibility of automated fetal heart rate measurement using a novel technique with two-dimensional tracking
Ryuhei Nagai1, Susumu Miyashita2, Susumu Murata3
1Department of Obstetrics and Gynecology, Kochi Health Sciences Center, Kochi, Japan.
Purpose:
To assess the accuracy of automated fetal heart rate measurement using two-dimensional tracking (AutoFHR) by comparison with the conventional free-angle M-mode (M-mode) and pulsed-waved Doppler (PWD) methods.
Methods:
A multicenter prospective comparative study was conducted. AutoFHR is a novel technique for the automatic calculation of fetal heart rate using a two-dimensional speckle-tracking method. The fetal heart rate (FHR) obtained by AutoFHR was compared with that obtained by the conventional M-mode and PWD. Statistical analysis was performed on the correlation between the FHR measured by AutoFHR and that determined by M-mode and PWD.
Results:
Data from 326 singleton pregnancies were analyzed, and all the data and the data from 178 cases were compared with M-mode and PWD, respectively. The intraobserver ICC was 0.96 (95% CI: 0.93-0.98), whereas the interobserver ICC was 0.97 (95% CI: 0.95-0.99). Systematic bias was not observed between M-mode and PWD, based on the Bland-Altman plots. Analyses of the relationships among the FHRs yielded by each method revealed that AutoFHR was strongly associated with M-mode (rs = 0.99, p < 0.001) and PWD (rs = 0.86, p < 0.001).
Conclusion:
Evaluation of FHR by AutoFHR was proved to be equivalent to evaluation by the conventional M-mode and PWD. AutoFHR can be employed with only B-mode data, making FHR measurement easier and safer.
Related Concept Videos
Measuring Reaction Rates
Regulation of Heart Rates
The SNS increases heart rate through the release of norepinephrine and epinephrine, which act on beta-1 adrenergic receptors in the heart. This action increases the rate of depolarization in the sinoatrial (SA) node, the heart's...
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Reaction Rate
The mathematical representation of the change in the concentration of reactants and products, over time, is the rate...
Fetal Circulation
Two umbilical arteries transport blood from the fetus to the placenta. At the placenta, the blood absorbs oxygen and nutrients while simultaneously eliminating waste products. This oxygen-enriched and nutrient-rich blood then returns to the fetus through one...