Robotic Cochlear Implantation for Direct Cochlear Access
Published on: June 16, 2022
The accuracy of image-based safety analysis for robotic cochlear implantation
Purpose:
To evaluate the accuracy and reliability of image-based safety analysis for robotic cochlear implantation (RCI) in an ex vivo assessment.
Methods:
The accuracy was evaluated in a study on 23 human temporal bones. For image analysis, a computer-assisted safety analysis based on intraoperative cone beam computed tomography was implemented. The method automatically segments the drill tunnel and predicts the distance between the tunnel and the facial nerve. In addition, the drilling error at the target is predicted. The predicted distances were compared with the actually drilled distances measured in postoperative high-resolution micro-computed tomography scans. The automatic method was compared to accuracies associated with a manual analysis of the image data.
Results:
The presented computerized image-based analysis enabled the proximity of the facial nerve to the drill trajectory to be predicted with an accuracy of 0.22 ± 0.15 mm and drilling error at the target to be predicted with an accuracy of 0.11 mm ± 0.08 during N = 19 RCI procedures. The manual assessment of facial nerve proximity was performed with an accuracy of 0.34 ± 0.20 mm by a trained clinical expert.
Conclusion:
The assessment of intraoperative CT-based imaging presents multiple benefits over alternative safety mechanisms including early detection and applicability even in cases of malformation of the mastoid. This work presents a computer-assisted approach to image analysis that enables procedure safety measurements to be reliably performed with superior accuracy to other proposed safety methodologies, at a safe distance from the facial nerve. Its application must, however, be considered in relation to associated costs (time, cost, irradiation) and the dependence of the measure on a reliable preoperative segmentation.
More Related Videos
03:58Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
Related Concept Videos
Improving Translational Accuracy
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision
Survey Safety
Accuracy and Precision
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...