Local speed of sound estimation in tissue using pulse-echo ultrasound: Model-based approach
Marko Jakovljevic1, Scott Hsieh2, Rehman Ali1
1Department of Radiology, Stanford School of Medicine, Stanford, California 94305, USA.
Abstract:
A model and method to accurately estimate the local speed of sound in tissue from pulse-echo ultrasound data is presented. The model relates the local speeds of sound along a wave propagation path to the average speed of sound over the path, and allows one to avoid bias in the sound-speed estimates that can result from overlying layers of subcutaneous fat and muscle tissue. Herein, the average speed of sound using the approach by Anderson and Trahey is measured, and then the authors solve the proposed model for the local sound-speed via gradient descent. The sound-speed estimator was tested in a series of simulation and ex vivo phantom experiments using two-layer media as a simple model of abdominal tissue. The bias of the local sound-speed estimates from the bottom layers is less than 6.2 m/s, while the bias of the matched Anderson's estimates is as high as 66 m/s. The local speed-of-sound estimates have higher standard deviation than the Anderson's estimates. When the mean local estimate is computed over a 5-by-5 mm region of interest, its standard deviation is reduced to less than 7 m/s.
Related Concept Videos
Speed of Sound in Gases
Deriving the Speed of Sound in a Liquid
The speed of sound in fluids can be derived by considering a mechanical wave...
Speed of Sound in Solids and Liquids
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Pulse
The pulse serves as a clinical...
Korotkoff Sounds
During blood pressure assessment, inflating the cuff 30 millimeters of mercury above the patient's systolic blood pressure...