Related Experiment Video
Updated: Feb 6, 2026

Performing Colonoscopic-Guided Pinch Biopsies in Mice and Evaluating Subsequent Tissue Changes
Published on: February 5, 2021
Faster colonoscope withdrawal time without impaired detection using EndoRings
John C Thygesen1, Prasanna Ponugoti1, William W Tippins1
1Division of Gastroenterology/Hepatology, Indiana University School of Medicine, Indianapolis, Indiana, United States.
EndoRings, a device for colonoscopy, may speed up the withdrawal phase without impacting adenoma detection rates. Further research is recommended for mucosal exposure devices to optimize procedure times.
Area of Science:
- Gastroenterology
- Endoscopy
- Medical Devices
Background:
- Mucosal exposure devices enhance polyp detection during colonoscopy.
- EndoRings are a type of mucosal exposure device evaluated in this study.
Purpose of the Study:
- To evaluate the impact of EndoRings on adenoma detection and procedure times in colonoscopies.
- To compare colonoscopies performed with and without EndoRings.
Main Methods:
- A university practice conducted a trial with 14 endoscopists using EndoRings.
- Colonoscopy procedures with EndoRings were compared to age- and indication-matched procedures without the device.
Main Results:
- The adenoma detection rate was 44% with EndoRings versus 39% without (P=0.39).
- Mean adenomas per colonoscopy were 1.2 with EndoRings versus 0.9 without (P=0.055).
- Mean withdrawal time was significantly faster with EndoRings (12.2 minutes) compared to conventional colonoscopy (16.1 minutes, P=0.0005).
Conclusions:
- EndoRings may facilitate faster withdrawal times during colonoscopy.
- The use of EndoRings did not compromise adenoma detection rates.
- Prospective trials focusing on procedure times are warranted for mucosal exposure devices.
More Related Videos
06:48Quantitative Real-Time Polymerase Chain Reaction Evaluation of MicroRNA Expression in Kidney and Serum of Mice with Age-Dependent Renal Impairment
Published on: April 29, 2022
05:26Author Spotlight: Innovations in iTUG Test for Enhanced Risk Assessment and Cognitive Insights
Published on: October 25, 2024
Related Concept Videos
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Effect of Hepatic Disease on Pharmacokinetics: Dose Adjustments Due to Hepatic Impairment
Mean free path and Mean free time
Mixing Time
Generation Time
Time Course of Drug Effect