[Comparison of different massive parallel sequencing platforms for mutation profiling in formalin-fixed and
R R Jiang1, Y J Wang, X D Teng
1Department of Pathology, West China Hospital, Sichuan University, Chengdu 610041, China.
The Miseq platform offers superior next-generation sequencing (NGS) performance for formalin-fixed and paraffin-embedded (FFPE) samples compared to Ion Torrent PGM. Miseq provides higher quality reads, alignment, and depth, making it more reliable for detecting mutations in FFPE cancer samples.
Area of Science:
- Molecular Biology
- Genomics
- Oncology
Background:
- Next-generation sequencing (NGS) is crucial for molecular characterization of tumors.
- Formalin-fixed and paraffin-embedded (FFPE) samples present unique challenges for NGS due to potential DNA degradation.
- Comparing different NGS platforms is essential for optimizing clinical diagnostics.
Purpose of the Study:
- To evaluate and compare the performance of Miseq and Ion Torrent PGM platforms for NGS.
- To assess the efficacy of these platforms in analyzing DNA from FFPE cancer samples.
- To determine the optimal NGS platform for reliable mutation detection in FFPE tissues.
Main Methods:
- Retrospective analysis of 204 FFPE cancer samples (100 non-small cell lung, 104 colorectal).
- DNA extraction and library preparation using identical kits and protocols for both platforms.
- Sequencing performed on Miseq and Ion Torrent PGM platforms, followed by validation of discordant mutations using ARMS-PCR and Sanger sequencing.
Main Results:
- Miseq platform generated significantly higher numbers of reads, superior alignment rates, and greater median sequence depth compared to PGM.
- A high concordance (93.6%) of detected mutations was observed between the two platforms.
- Miseq demonstrated higher reliability for detecting low-frequency mutations, with discordant results often attributed to low mutation frequency or sample age.
Conclusions:
- Miseq platform exhibits superior sequencing quality and reliability for analyzing FFPE samples over Ion Torrent PGM.
- The choice of NGS platform should consider sample characteristics, throughput needs, and desired reliability for molecular tumor profiling.
- Miseq is recommended for more dependable molecular characterization of FFPE cancer specimens.