Thermal Ablation for the Treatment of Abdominal Tumors
Published on: March 7, 2011
Overlapping radiofrequency ablation planning and robot-assisted needle insertion for large liver tumors
Background:
We aimed at facilitating percutaneous radiofrequency ablation (RFA) for large tumors with accurate overlapping ablation planning and robot-assisted needle insertion from a single incision port (SIP).
Methods:
We developed a personalized and quantitative RFA planning method to obtain multiple needle overlapping ablation planning through a single incision. A robot with a remote center of motion mechanism was designed to perform needle insertions through a SIP according to the planning.
Results:
Numerical and visual evaluation showed that the planning could yield nearly full coverage over large tumors and greatly reduce both ablation times and the ablations of normal tissues. Ex vivo and in vivo experiments showed that our robot-assisted needle insertion system was capable of conducting the RFA procedure in large liver tumors from a SIP.
Conclusions:
Robot-assisted RFA needle insertions from a SIP makes RFA treatment for large tumors more minimally invasive, predictable, and repeatable.
Related Concept Videos
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...
Sampling Plans
Random sampling is a method where each member of the population has an equal chance of being selected for the sample. It involves selecting individuals randomly, often using random number generators or lottery-type methods. For example, when analyzing the properties of a...
Planning Nursing Care I
Planning Nursing Care II
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Liver Histology
Hepatocytes perform a variety of essential functions. They secrete...