Related Experiment Video
Updated: Feb 6, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Long-term use of the wearable cardioverter defibrillator in patients with explanted ICD
Georgy Kaspar1, Kumar Sanam1, Gunjan Gholkar1
1Department of Cardiology, Providence-Providence Park Hospital/Michigan State University College of Human Medicine, Southfield, MI, USA.
Objectives:
To evaluate the effectiveness of wearable cardioverter defibrillator (WCD) use in protecting patients from sudden cardiac arrest (SCA) while they were treated in nonhospital settings until re-implantation of an Implantable cardioverter-defibrillator (ICD) was feasible. We sought to determine whether the WCD could be successfully utilized long term (≥1 year) after ICD extraction in patients at continued risk of SCD in which ICD re-implantation was not practical.
Background:
ICDs have proven to improve mortality in patients for both secondary and primary prevention of SCA. Increased ICD implantation in older patients with comorbid conditions has resulted in higher rates of cardiac device infections. Currently, a wearable cardioverter defibrillator (WCD) is an alternative management for SCA prevention in specific cases.
Methods:
This a retrospective analysis based on consecutive WCD patients who underwent ICD explant due to device-related infections or mechanical reasons between April 2007 and July 2014. A total of 102 patients were identified from the national database maintained by ZOLL (Pittsburgh, PA, USA). We analyzed the reason for WCD use, demographic information, device data, compliance and duration of WCD use, detected arrhythmias and therapies, and reason for discontinuing WCD use.
Results:
In these long term WCD users, average length of WCD use was 638 ± 361 days. Nine patients (8.8%) had a sustained ventricular arrhythmia that was successfully resuscitated by the WCD. Six patients (5.8%) experienced inappropriate shocks. Two patients (1.9%) died of asystole events while wearing the WCD and an additional 10 patients died while not monitored by the WCD. Thirty-nine patients (38.2%) ended WCD use when a new ICD was implanted and 15 patients (14.7%) were still wearing the WCD at the time of analysis.
Conclusions:
We found that extending use of the WCD to ≥1 year is a safe and effective alternative treatment for patients with explanted ICDs who are not pacemaker dependent.
More Related Videos
Related Concept Videos
Long-term Depression
Long-term Depression
Calcium Ion Concentration Mechanism
If over...
Long-term Potentiation
Long-term Potentiation
Hebbian LTP
LTP can occur when...
Directional Terms
Regional Terms
Primarily, the human body has two major regions, the axial and appendicular regions. The axial region comprises regions from the head to the abdomen and makes up the central body axis. In contrast,...