A novel rectal applicator for contact radiotherapy with HDR 192Ir sources
Murillo Bellezzo1, Gabriel P Fonseca2, An-Sofie Verrijssen2
1Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht, The Netherlands; Centro de Engenharia Nuclear, Instituto de Pesquisas Energéticas e Nucleares IPEN-CNEN/SP, São Paulo, Brazil.
Purpose:
Dose escalation to rectal tumors leads to higher complete response rates and may thereby enable omission of surgery. Important advantages of endoluminal boosting techniques include the possibility to apply a more selective/localized boost than using external beam radiotherapy. A novel brachytherapy (BT) rectal applicator with lateral shielding was designed to be used with a rectoscope for eye-guided positioning to deliver a dose distribution similar to the one of contact x-ray radiotherapy devices, using commonly available high-dose-rate 192Ir BT sources.
Methods And Materials:
A cylindrical multichannel BT applicator with lateral shielding was designed by Monte Carlo modeling, validated experimentally with film dosimetry and compared with results found in the literature for the Papillon 50 (P50) contact x-ray radiotherapy device regarding rectoscope dimensions, radiation beam shape, dose fall-off, and treatment time.
Results:
The multichannel applicator designed is able to deliver 30 Gy under 13 min with a 20350 U (5 Ci) source. The use of multiple channels and lateral shielding provide a uniform circular treatment surface with 22 mm in diameter. The resulting dose fall-off is slightly steeper (maximum difference of 5%) than the one generated by the P50 device with the 22 mm applicator.
Conclusions:
A novel multichannel rectal applicator for contact radiotherapy with high-dose-rate 192Ir sources that can be integrated with commercially available treatment planning systems was designed to produce a dose distribution similar to the one obtained by the P50 device.
Related Concept Videos
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Contact-dependent Signaling
Gap Junctions
In animal cells, gap junctions are formed...
Sinusoidal Sources
In homes, the power supplies use sinusoidal sources to provide electricity. These sources generate a voltage that varies sinusoidally...
AC Sources
Sources of Law
Constitutional law is foundational, deriving from federal and state constitutions, and...