Fluid Dynamics in the HeartMate 3: Influence of the Artificial Pulse Feature and Residual Cardiac Pulsation
Lena Wiegmann1, Bente Thamsen2,3,4, Diane de Zélicourt1,3
1The Interface Group, Institute of Physiology, University of Zurich, Zurich, Switzerland.
Insights
The HeartMate 3 ventricular assist device (VAD) shows good flow, but its artificial pulse increases turbulence and total stresses. This may explain hemocompatibility issues despite low pump thrombosis rates.
Area of Science:
- Biomedical Engineering
- Fluid Dynamics
- Cardiovascular Science
Background:
- Ventricular assist devices (VADs) are crucial for end-stage heart failure patients.
- The HeartMate 3 (HM3) VAD features an artificial pulse to mitigate thrombosis.
- Stroke and bleeding complications persist despite VAD advancements.
Purpose of the Study:
- To investigate the impact of HM3's artificial pulse on internal flow dynamics.
- To analyze flow fields and stresses under various operational modes.
- To correlate flow characteristics with clinical complications like hemocompatibility issues.
Main Methods:
- High-resolution computational fluid dynamics (CFD) simulations were employed.
- Eulerian and Lagrangian flow features were compared.
- Simulations included constant speed, artificial pulse, and native cardiac cycle effects.
Main Results:
- Good scalar washout was observed across all tested conditions, correlating with low pump thrombosis.
- The artificial pulse did not improve scalar washout but increased velocity and gradient variations.
- Viscous stresses were lower than in other VADs, but the pulse significantly increased turbulence and total stresses.
Conclusions:
- The HM3 VAD demonstrates effective washout, potentially explaining low thrombosis rates.
- The artificial pulse's impact on turbulence and stresses may be linked to hemocompatibility issues.
- Further research is needed to fully understand the clinical implications of HM3's flow dynamics.
Abstract:
Ventricular assist devices (VADs), among which the HeartMate 3 (HM3) is the latest clinically approved representative, are often the therapy of choice for patients with end-stage heart failure. Despite advances in the prevention of pump thrombosis, rates of stroke and bleeding remain high. These complications are attributed to the flow field within the VAD, among other factors. One of the HM3's characteristic features is an artificial pulse that changes the rotor speed periodically by 4000 rpm, which is meant to reduce zones of recirculation and stasis. In this study, we investigated the effect of this speed modulation on the flow fields and stresses using high-resolution computational fluid dynamics. To this end, we compared Eulerian and Lagrangian features of the flow fields during constant pump operation, during operation with the artificial pulse feature, and with the effect of the residual native cardiac cycle. We observed good washout in all investigated situations, which may explain the low incidence rates of pump thrombosis. The artificial pulse had no additional benefit on scalar washout performance, but it induced rapid variations in the flow velocity and its gradients. This may be relevant for the removal of deposits in the pump. Overall, we found that viscous stresses in the HM3 were lower than in other current VADs. However, the artificial pulse substantially increased turbulence, and thereby also total stresses, which may contribute to clinically observed issues related to hemocompatibility.
Related Concept Videos
Residual Plots
When the residual values are plotted against the variable x, it is called a residual...
Residual Stresses
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Residual Stresses in Circular Shafts
Residual Stresses in Bending