Orthopedic Robot-Assisted Femoral Neck System in the Treatment of Femoral Neck Fracture
Published on: March 3, 2023
Combined application of virtual monoenergetic high keV images and the orthopedic metal artifact reduction algorithm
Junghoan Park1,2, Se Hyung Kim3,4, Joon Koo Han1,2,5
1Department of Radiology, Seoul National University Hospital, 101 Daehak-ro, Jongno-gu, Seoul, 03080, Republic of Korea.
Purpose:
To determine whether there is any additional metal artifact reduction when virtual monochromatic images (VMI) and metal artifact reduction for orthopedic implants (O-MAR) are applied together compared to their separate application in both phantom and clinical abdominopelvic CT studies.
Methods:
An agar phantom containing a spinal prosthesis was scanned using a dual-layer, energy CT scanner (IQon, Philips Healthcare), and reconstructed with the filtered back-projection algorithm without O-MAR (FBP), filtered back-projection algorithm with O-MAR (O-MAR), VMI140 without O-MAR (VMI140), and VMI140 with O-MAR (VMI140 + O-MAR). Abdominopelvic CT images of 47 patients with metallic prostheses were also reconstructed in the same manner for clinical study. Noise measured as the standard deviation of CT Hounsfield units was compared between the four reconstruction methods in both phantom and clinical studies. Improvements in metal artifact reduction, image quality, and diagnostic improvement were further analyzed in the clinical study.
Results:
Noise was significantly decreased when both VMI and O-MAR were applied in conjunction compared to their separate application in both phantom (16.3 HU vs. 25.0 and 26.4 HU) and clinical studies (15.8 HU vs. 19.2 and 26.2 HU). In the clinical study, the qualitative degree of artifacts was also significantly reduced with VMI140 + O-MAR (2.85 and 2.87) compared to VMI140 (2.36 and 2.26) or O-MAR (2.13 and 2.04) alone for both reviewers (P < 0.001) and improvements in image quality were observed in all 47 patients, with actual diagnostic improvements observed in three.
Conclusions:
Metal artifacts can be additionally reduced by applying O-MAR and VMI in conjunction, compared to their separate application, thereby improving diagnostic performance.
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Oxidation-Reduction Reactions
Metal-Ligand Bonds
In these complexes, transition metals form coordinate covalent bonds, a kind of Lewis acid-base interaction in which both of the electrons in the bond are contributed by a donor (Lewis base) to an electron acceptor (Lewis acid). The Lewis acid in...
Virtual Work
In static equilibrium, a body can experience an imaginary or virtual movement, such as displacement or rotation. The virtual work done by a force is equal to the dot product of force and virtual displacement in the direction of the force. When it comes to virtually rotating a...
Bonding in Metals
Metallic Solids
All metallic solids exhibit high thermal and electrical conductivity, metallic luster, and malleability....