Filiz Uzumcugil1, Emre Can Celebioglu2, Demet Basak Ozkaragoz1
1Department of Anesthesiology and Reanimation, Hacettepe University School of Medicine, Ankara, Turkey.
Objectives:
The age-based Cole formula has been employed for the estimation of endotracheal tube (ETT) size due to its ease of use, but may not appropriately consider growth rates among children. Child growth is assessed by calculating the body surface area (BSA). The association between the outer diameter of an appropriate uncuffed-endotrachealtube (ETT-OD) and the BSA values of patients at 24-96 months of age was our primary outcome.
Methods:
Cole formula, BSA, age, height, weight and ultrasound measurement of subglottic-transverse-diameter were evaluated for correlations with correct uncuffed ETT-OD. The Cole formula, BSA, and ultrasound measurements were analyzed for estimation rates in all patients and age subgroups. The maximum allowed error for the estimation of ETT-OD was ≤0.3 mm. Patients' tracheas were intubated with tubes chosen by Cole formula and correct ETT-OD values were determined using leak test. ETT exchange rates were recorded.
Results:
One-hundred twenty-seven patients were analyzed for the determination of estimation rates. Thirteen patients aged ≥72 months were intubated with cuffed ETT-OD of 8.4 mm and were accepted to need uncuffed ETT-OD >8.4 mm in order to be included in estimation rates, but excluded from correlations for size analysis. One-hundred fourteen patients were analyzed for correlations between correct ETT-OD (determined by the leak test) and outcome parameters. Cole formula, ultrasonography, and BSA had similar correct estimation rates. All three parameters had higher underestimation rates as age increased.
Conclusion:
The Cole formula, BSA, and ultrasonography had similar estimation rates in patients aged 24≤ to ≤96 months. BSA had a correct estimation rate of 40.2% and may not be reliable in clinical practice to predict uncuffedETT-size.
Related Concept Videos
Reliability and Validity
Factors Affecting Dissolution: Particle Size and Effective Surface Area
Cell Size
Surface Area
Cells can take in nutrients and water via diffusion through the plasma membrane itself or through specific channels in the membrane. The area of the membrane surrounding...
Distribution Reliability and Automation
Uterine Tubes
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....