Validation of frame-based positioning accuracy with cone-beam computed tomography in Gamma Knife Icon radiosurgery
Florian Stieler1, Frederik Wenz1, Beate Schweizer1
1Department of Radiation Oncology, Universitätsmedizin Mannheim, University of Heidelberg, Mannheim, Germany.
Background:
Frame based positioning accuracy in Gamma Knife (GK) stereotactic radiosurgery (SRS) is extremely high but removal of a post may be necessary to enable the treatment in selected patients.
Objective:
To verify the positioning accuracy in clinical scenarios with 4 and 3 posts in patients and phantoms using cone-beam CT (CBCT) of Gamma Knife Icon™.
Methods:
We analyzed positioning accuracy for 12 patients with standard 4 post setup using pretreatment CBCT (pre-CBCT) on GK Icon™ and report 4 patients with different clinical scenarios (removal of a post). We performed phantom measurements to verify the frame accuracy via CBCT in different clinical scenarios without the influence of the human patient.
Results:
Mean frame accuracy for 12 patients with 4 posts was 0.35 mm/0.34 degree. Mean motion during treatment was 0.11 mm/0.04 degree. For two of the clinical scenarios where a post was removed, we found acceptable deviations within 0.66 mm/0.61 degree. For 2 patients, a deviation of 2.94 mm/-3.47 degree and 1.85 mm/-0.74 degree was found and replanning was necessary. Phantom measurements showed good agreement when planning MR/CT was performed with 4 or 3 post. Larger deviations of 0.86 mm/0.88 degree were detected when a post was removed after planning MR/CT.
Conclusion:
The frame accuracy with 4 posts before and during GK treatments is as high as expected. For clinical situations, where a post is removed after planning-CT/MR, pre-treatment position verification is strongly suggested using stereotactic CBCT or the P-CT/MR should be repeated to avoid possible mistreatments.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Framing Effects
Imaging Studies III: Computed Tomography
Reliability and Validity
Improving Translational Accuracy
Frames
Frames are versatile and widely used in various applications such as structural supports for beams and columns, automobile chassis construction, and in the construction...