Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Validation of new transmission detector transmission factors for online dosimetry: an experimental study
So-Yeon Park1,2, Jong Min Park2,3,4, Jung-In Kim2,3,4
1Department of Radiation Oncology, Veterans Health Service Medical Center, Seoul, Republic of Korea.
Background:
The demand for dose verification during treatment has risen with the increasing use of intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) in modern radiation therapy. This study aims to validate the transmission factors of a new transmission detector, the Dolphin online monitoring system (IBA Dosimetry, Schwarzenbruck, Germany), for clinical use.
Methods:
The transmission factors of the Dolphin detector were evaluated using 6 MV, 6 flattening filter free (FFF), 10 MV, and 10 FFF clinical beams from a TrueBeam STx linear accelerator system. Two-dimensional (2D) dose distributions were measured through portal dosimetry with and without Dolphin to derive the transmission factors. The measurements were performed using 10 IMRT and 10 VMAT treatment plans. The transmission factors were calculated using a non-negative least squares problem solver for the 2D dose matrix. Normalized plans were generated using the derived transmission factors. Patient-specific quality assurance with normalized plans was performed using portal dosimetry and an ArcCheck detector to verify the transmission factors. The gamma passing rates were calculated for the 2%/2 mm and 1%/1 mm criteria.
Results:
The transmission factors for the 6 MV, 6 FFF, 10 MV, and 10 FFF beams, were 0.878, 0.824, 0.913, and 0.883, respectively. The average dose difference between the original plan without Dolphin and the normalized plan with Dolphin was less than 1.8% for all measurements. The mean passing rates of the gamma evaluation were 98.1 ± 2.1 and 82.9 ± 12.6 for the 2%/2 mm and 1%/1 mm criteria, respectively, for portal dosimetry of the original plan. In the case of the portal dosimetry of the normalized plan, the mean passing rates of the gamma evaluation were 97.2 ± 2.8 and 79.1 ± 14.8 for the 2%/2 mm and 1%/1 mm criteria, respectively.
Conclusions:
The Dolphin detector can be used for online dosimetry when valid transmission factors are applied to the clinical plan.
More Related Videos
09:02An Experimental Model to Study Tuberculosis-Malaria Coinfection upon Natural Transmission of Mycobacterium tuberculosis and Plasmodium berghei
Published on: February 17, 2014
10:29Studying Dynamic Processes of Nano-sized Objects in Liquid using Scanning Transmission Electron Microscopy
Published on: February 5, 2017
Related Concept Videos
Design of Transmission Shafts
Transmission Line Design Considerations
Transmission Shafts: Problem Solving
Next, use bending moment diagrams for the shaft to...
Transmission-Line Differential Equations
Line Section Model
A circuit representing a line section of length Δx helps in understanding the transmission line parameters. The voltage V(x) and current i(x) are measured from...
Transmission Electron Microscopy
Reliability and Validity