Correlation between invasive and noninvasive blood pressure measurements in severely burned children
Janos Cambiaso-Daniel1, Victoria G Rontoyanni2, Guillermo Foncerrada3
1Department of Surgery, University of Texas Medical Branch and Shriners Hospitals for Children, Galveston, TX USA; Division of Plastic, Aesthetic and Reconstructive Surgery, Department of Surgery, Medical University of Graz, Austria.
Introduction:
Accurate blood pressure monitoring is essential for burn management, with the intra-arterial line method being the gold standard. Here we evaluated agreement between cuff and intra-arterial line methods.
Methods:
Data from burned children admitted from 1997 to 2016 were retrospectively reviewed. Simultaneously collected intra-arterial and cuff measurements were cross-matched and linear regression performed to assess agreement for systolic blood pressure (SBP), diastolic blood pressure (DBP), and mean arterial pressure (MAP).
Results:
We identified 9969 matches for SBP, DBP, and MAP in 872 patients (579 male) aged 8±5years with burns covering 52±20% of the total body surface area and a hospitalization lasting 33±31 days. Intra-arterial lines had a complication rate of 1%. The mean bias (95% CI) between methods was 1.3 (0.5, 2.1) mm Hg for SBP, -6.4 (-7.0, -5.7) mmHg for DBP, and -5.8 (-6.4, -5.3) mmHg for MAP. The standard deviation of the bias (95% limit of agreement) was 12.1 (-22.5, 25.1) mmHg for SBP, 9.9 (-25.8, 13.0) mmHg for DBP, and 8.7 (-22.8, 11.1) mmHg for MAP.
Conclusions:
Cuff measurements vary widely from those of intra-arterial lines, which have a low complication rate. Intra-arterial lines are advisable when tight control of the hemodynamic response is essential.
Related Concept Videos
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...