Mechanical performance of a commercial knowledge-based VMAT planning for prostate cancer
Mikoto Tamura1, Hajime Monzen2, Kenji Matsumoto1
1Department of Medical Physics, Graduate School of Medical Science, Kindai University, 377-2, Ohno-Higashi, Osakasayama, Osaka, 589-8511, Japan.
Background:
This study clarified the mechanical performance of volumetric modulated arc therapy (VMAT) plans for prostate cancer generated with a commercial knowledge-based treatment planning (KBP) and whether KBP system could be applied clinically without any major problems with mechanical performance.
Methods:
Thirty consecutive prostate cancer patients who underwent VMAT using extant clinical plans were evaluated. The mechanical performance and dosimetric accuracy of the single optimized KBPs, which were trained with other 51 clinical plans, were compared with the clinical plans. The mechanical performance metrics were mean field area (MFA), mean asymmetry distance (MAD), cross-axis score (CAS), closed leaf score (CLS), small aperture score (SAS), leaf travel (LT), modulation complexity score (MCSv), and monitor unit (MU). The γ passing rates were evaluated with ArcCheck and EBT3 film.
Results:
The mean mechanical performance metrics (clinical plan vs. KBP) were as follows: 18.28 cm2 vs. 17.25 cm2 (MFA), 21.08 mm vs. 20.47 mm (MAD), 0.54 vs. 0.55 (CAS), 0.040 vs. 0.051 (CLS), 0.20 vs. 0.23 (SAS5mm), 458.5 mm vs. 418.8 mm (LT), 0.27 vs. 0.27 (MCSv), and 618.2 vs. 622.1 (MU), respectively. Significant differences were observed for CLS and LT. The average γ passing rates (clinical plan vs. KBP) were as follows: 99.0% vs. 99.1% (3%/3 mm) and 92.4% vs. 92.5% (2%/2 mm) with ArcCHeck, and 99.5% vs. 99.4% (3%/3 mm) and 95.2% vs. 95.4% (2%/2 mm) with EBT3 film, respectively.
Conclusions:
The KBP used lower multileaf collimator (MLC) travel and more closed or small MLC apertures than the clinical plan. The KBP system of VMAT for the prostate cancer was acceptable for clinical use without any major problems.
Related Concept Videos
Adaptive Mechanisms in Cancer Cells
Some of the advantages that cancer cells have on normal cells include - enhanced ability to divide without terminally differentiating, induce new blood vessel formation,...
Mechanisms of Retrovirus-induced Cancers
Sampling Plans
Random sampling is a method where each member of the population has an equal chance of being selected for the sample. It involves selecting individuals randomly, often using random number generators or lottery-type methods. For example, when analyzing the properties of a...
Planning Nursing Care I
Planning Nursing Care II
Mechanical Protein Functions