Published on: June 16, 2022
Effectiveness of skull X-RAY to determine cochlear implant insertion depth
Vinay Fernandes1, Yiqiao Wang2, Robert Yeung3
1Division of Otolaryngology - Head and Neck Surgery, Faculty of Medicine, University of Toronto, Toronto, Canada.
Background:
Cochlear implant (CI) insertion depth can affect residual hearing preservation, tonotopic range coverage, and Mapping. Therefore, determining insertion depth has the potential to maximize CI performance. A post-op skull X-RAY is commonly used to assess insertion depth, however its effectiveness has not been well established. Our primary objective was to assess the accuracy of post-op skull X-RAYs to determine insertion depth, compared to CT as the gold standard. Secondary objectives were to compare experience level of raters and different skull X-RAY views.
Methods:
Thirteen patients with Advanced Bionic HiRes 90 K implants, and post-operative temporal bone CT scans were selected from the CI database at Sunnybrook Health Sciences Centre. Medical students, otology fellows, and CI surgeons evaluated insertion depths on post-op skull X-RAYs, while neuroradiologists evaluated CT scans. Descriptive statistics, regression analysis, and paired t-tests were used to compare the two types of imaging.
Results:
X-RAYs and CTs provided an equivalent mean insertion depth of 337 degrees (p = 0.93), a mean difference of - 0.9 degrees and a standard deviation of paired differences of 43 degrees. Although means were similar across rater groups, CI surgeons (45 degrees) had the lowest standard deviation of paired differences. Comparing X-RAY views, Caldwell (29 degrees) had less variation than Towne (59 degrees) for standard deviation of paired differences.
Conclusions:
Skull X-RAYs provide accurate and reliable measurements for CI insertion depth. The Caldwell view alone may be sufficient for evaluations of insertion depth, and experience has a minor impact on the variability of estimates.
Related Concept Videos
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Sutures of the Skull
Sutures are immobile joints between adjacent bones of the skull. The narrow gap between the bones is filled with dense, fibrous connective tissue that unites the bones. The long sutures located between the skull bones are not straight but instead follow irregular, tightly twisting paths. These twisting lines tightly...
Biological Effects of Radiation
Overview of the Skull
The cranial vault surrounds and protects the brain and houses the middle and inner ear structures. This cavity is bounded superiorly by the rounded top of the skull, which...
Uniform Depth Channel Flow
Depth Perception and Spatial Vision