Usefulness of the Impulse Oscillometry System in Graft Function Monitoring in Lung Transplant Recipients
M Ochman1, J Wojarski2, A Wiórek2
1Department of Cardiac, Vascular and Endovascular Surgery and Transplantology, Medical University of Silesia in Katowice, Silesian Centre for Heart Diseases, Zabrze, Poland; Department of Pharmacology, School of Medicine with the Division of Dentistry in Zabrze, Zabrze, Poland.
Background:
After successful lung transplantation, patients are monitored for chronic lung allograft dysfunction. Pulmonary function tests and 6-minute-walk tests are commonly used for functional graft monitoring. As these methods require substantial effort, however, many patients are unable to complete testing fully. The impulse oscillometry system is a noninvasive method that requires minimal patient cooperation and is suitable for use for patients incapable of strenuous activity. We compared impulse oscillometry system with pulmonary function tests and 6-minute-walk tests to determine if impulse oscillometry system could serve as a substitute measure.
Methods:
This prospective, observational study evaluated 25 consecutive patients (19 men, median age 54.5 years) admitted to a single institution from January to October 2016 (double-lung = 13, single-lung = 13). Patients were assessed using pulmonary function tests, impulse oscillometry system, and 6-minute-walk tests.
Results:
Eighty-eight percent of patients reached high-resonance frequency (Fres) and, in 84% of patients, the value of Ax (area of reactance) increased above the norm (N < 0.33 kPa/L) indicating peripheral airways obstruction. High resistance of small airways, measured with an R5 - R20 difference, followed higher Ax values. The increase of resistance at 5 Hz in 31% of patients (R5 >150% predicted value) also indicated small airway obstruction. Airway obturation in patients with elevated Ax and R5 was confirmed by decreased FEV1 (<75% predictive value) and FEV1/FVC ratio in 38% of patients.
Conclusions:
Study results confirm the impulse oscillometry system method could be a substitute for pulmonary function tests in determining the occurrence of chronic lung allograft dysfunction. 6-minute-walk tests showed neither strong correlations regarding impulse oscillometry system and pulmonary function tests nor any base for differentiation of results regarding main factor codes.
Related Concept Videos
Impulse
Additionally, it can be shown that the...
Impulse Response
Kirchhoff's law forms an input signal equation, with the capacitor's current and voltage providing the output. Substituting the current and dividing by RC yields a differential equation. The output for an impulse input is the impulse...
Lung Capacity
Principle of Impulse and Moment
Impulse-Momentum Theorem
By writing Newton's second law of motion in terms of the momentum of an object and the external force acting on it, and simultaneously using the definition of the impulse vector, it can be shown that the total impulse on an object is equal to its...
Principle of Angular Impulse and Momentum: Problem Solving
Initially, a free-body diagram of the system is drawn to illustrate all the forces acting upon the system, providing a crucial understanding of the dynamics at play. Then, the principle of angular impulse and momentum is...