Related Experiment Video
Updated: Feb 5, 2026

Fabrication of Small Caliber Stent-grafts Using Electrospinning and Balloon Expandable Bare Metal Stents
Published on: October 26, 2016
Severe infolding of fenestrated-branched endovascular stent graft
Aleem K Mirza1, Giuliano A Sandri1, Emanuel R Tenorio1
1Division of Vascular and Endovascular Surgery, Mayo Clinic, Rochester, Minn.
Abstract:
Infolding of a fenestrated-branched stent graft is an infrequent complication due to excessive oversizing. We report the case of an 89-year-old man who underwent a four-vessel fenestrated-branched endovascular aortic repair for a pararenal aortic aneurysm. Computed tomography angiography revealed severe infolding across the mesenteric-renal vessels. The patient was treated by angioplasty and placement of Palmaz stent. Cone-beam computed tomography confirmed patent visceral vessels with resolution of the infolding. This case illustrates an uncommon complication that can be prevented by modifications in the stent design and by immediate assessment using intraoperative cone-beam computed tomography.