Novel extraction technique of retained pacemaker and defibrillator lead during heart transplantation
Eriko Hasumi1,2, Katsuhito Fujiu2,3, Toshiya Kojima2
1Department of Ubiquitous Health Informatics, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan.
Background:
Removal of cardiac implantable electronic devices (CIEDs) by manual traction during orthotopic heart transplantation (OHT) sometimes results in retained lead fragments. Moreover, abandoned leads and retained lead fragments are a contraindication for magnetic resonance imaging (MRI) and may be a cause of CIED infection.
Objective:
To eliminate complications of retained lead fragments, we completely removed residual leads using an excimer laser sheath technique during OHT. We report our clinical experience and high success rate of lead extraction using the excimer laser sheath compared with manual traction during OHT.
Methods And Results:
We obtained data on 84 consecutive patients receiving OHT between August 2007 and August 2017. Thirty-nine of 84 patients had undergone CIED implantation before OHT and removal of all their leads was attempted during OHT. From 2007 to 2014, defibrillator and pacemaker leads were extracted by manual traction in all patients (N = 22). After 2015, all leads were extracted with the excimer laser sheath, and surgical assistance was prepared for the procedure (N = 17). Complete procedural success was achieved in 100% of patients in the excimer laser group and 77% of patients in the manual traction group.
Conclusion:
Extraction of abandoned leads using the excimer laser sheath system during OHT is novel and safe technique, and has a higher success rate than extraction using manual traction during OHT.
Related Concept Videos
Anatomy of the Heart
Anatomy of the Heart
The heart has three layers: the innermost endocardium, the muscular myocardium, and the outer epicardium, all working together for optimal cardiac function.
Chambers of the Heart
The heart is made up of four...
Phase-lead and Phase-lag Controllers
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...
Overview of the Heart
The heart's structure...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...