Intraocular Pressure Measurement Accuracy and Repeatability of a Modified Goldmann Prism: Multicenter Randomized
Sean McCafferty1, Kyle Tetrault2, Ann McColgin3
1Arizona Eye Consultants, Intuor Technologies, Tucson, Arizona, USA.
Purpose:
To clinically evaluate a modified surface Goldmann applanation tomometer (GAT) prism for intraocular pressure (IOP) accuracy, repeatability, and safety.
Design:
Prospective, open-label, randomized, controlled, multicenter reference device reliability and validity analysis.
Methods:
A GAT and a modified surface GAT prism measured IOP on 173 unique eyes. The study design and analysis complied with FDA 510(k) and ANSIZ80.10-2014 guidelines. All eyes were randomized to IOP measurement by 1 of 5 standard prisms or 5 modified prisms, each from a different manufacturing lot. Pressures were measured by 6 investigators, 2 times with each prism, for a total of 1384 IOP measurements. Analysis included Bland-Altman difference accuracy, intraoperator and interoperator IOP measurement, and manufactured lot repeatability.
Results:
Bland-Altman indicated no IOP measurements pairs outside the ±5 mm Hg guidelines. Operator and manufactured lot repeatability F tests and 1-way ANOVAs indicated no statistical difference between the standard and modified prisms (all P > .10). The difference in IOP measurements of the standard and modified prisms correlated well to Dresdner central corneal thickness (CCT) correction (P = .01).
Conclusion:
A modified surface replacement prism is statistically equivalent to a flat-surfaced prism. The modified surface prism indicated statistically significant correction for CCT requiring further testing outside the ANSI standard limits (0.500 mm < CCT < 0.600 mm) to examine its full potential.
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Clinical Trials
There are four phases in a clinical trial. A phase one...
Clinical Trials: Overview
Statistical Software for Data Analysis and Clinical Trials
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...