Effect of Nebulizer Location and Spontaneous Breathing on Aerosol Delivery During Airway Pressure Release Ventilation
Hui-Qing Ge1, Ji-Mei Wang1, Hui-Ling Lin2
11 Department of Respiratory Care, Sir Run Run Shaw Hospital, School of Medicine, Zhejiang University, Hangzhou, China.
Background:
Airway pressure release ventilation (APRV) maintains a sustained airway pressure over a large proportion of the respiratory cycle, and has a long inspiratory time at high pressure. The purpose of this study was to determine the influence of the APRV with and without spontaneous breathing on albuterol aerosol delivery with a continuous vibrating-mesh nebulizer (VMN) placed at different positions on an adult lung model of invasive mechanical ventilation.
Methods:
An adult lung model was assembled by connecting a ventilator with a dual-limb circuit to an 8-mm inner diameter endotracheal tube (ETT) and collecting filter attached to a test lung with set compliance of 0.1 L/cmH2O and resistance of 0.5 cmH2O/(L·s). Four ventilator modes were compared: pressure control ventilation (PCV) with no bias flow, PCV with bias flow of 6 L/min (PCVBF6), APRV with no spontaneous breaths (APRV), and APRV with spontaneous breath trigger (APRVs). Peak inspiratory pressure, peak end-expiratory pressure, aerosol dose, and nebulization time were similar for all modes. The VMN was placed (1) between Y-piece and inspiratory limb, (2) at the gas outlet of a heated humidifier, and (3) at the gas inlet of a heated humidifier. Albuterol sulfate (5 mg/2.5 mL) was administered with each run and collected on a filter distal to the ETT. Deposited drug was eluted from each filter (purified water) and analyzed by UV spectrophotometry at 276 nm. Analysis of variance [general linear model (GLM) multivariate] was performed using the linear model of multiple variables, significance at p < 0.05.
Results:
Albuterol (in micrograms, mean ± standard deviation) delivered was higher with VMN placed at the gas inlet of the humidifier with each mode of ventilation (p < 0.01). APRVs has the highest albuterol delivery followed by PCV, PCVBF6, and APRV (1706.2 ± 60.9 μg vs. 1490.6 ± 61.1 μg vs. 1182.3 ± 61.4 μg vs. 1153.1 ± 99.7 μg, respectively, p < 0.001). The minute volume was positively correlated with the inhaled albuterol dose.
Conclusions:
Spontaneous breathing increased the albuterol delivery during APRV, compared with APRV alone and PCV modes. Placing the nebulizer proximal to the ventilator was more efficient for all modes tested.
More Related Videos
15:04Modeling and Simulations of Olfactory Drug Delivery with Passive and Active Controls of Nasally Inhaled Pharmaceutical Aerosols
Published on: May 20, 2016
05:46A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
Related Concept Videos
Breathing
Spontaneity
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
In Vitro Drug Release Testing: Overview, Development and Validation
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Location and Orientation of the Heart