Related Experiment Video
Updated: Feb 5, 2026

Murine Model for Non-invasive Imaging to Detect and Monitor Ovarian Cancer Recurrence
Published on: November 2, 2014
The Woven EndoBridge (WEB) for recurrent aneurysms: Clinical and imaging results
Sbt van Rooij1, W J van Rooij2, M Sluzewski2
11 Department of Radiology, Noordwest Ziekenhuisgroep, Alkmaar, The Netherlands.
Introduction:
The Woven EndoBridge (WEB) device is a novel intrasaccular flow disrupter designed for wide-necked aneurysms. We present our results of WEB treatment in previously treated and reopened aneurysms.
Materials And Methods:
Between February 2015 and December 2017, 17 patients with reopening of previously treated aneurysms were treated using the WEB device. Initial treatment was clipping in one, WEB in five and coiling in 11 aneurysms. Six aneurysms had two or three previous treatments. Mechanism of aneurysm reopening was clip remnant in one, compaction without regrowth in three, focal regrowth in five, reopening in partially thrombosed aneurysms in three and WEB compression in five aneurysms.
Results:
Endovascular treatment of the recurrent aneurysm was possible in all 17 patients with good WEB position obtained. Additional devices were used in three of 17 patients (18%): coils in two and stent in one patient. No technical, hemorrhagic or thromboembolic complications occurred (0%, 97.5% confidence interval 0-22%). Overall imaging result at latest follow-up was complete occlusion in five (29%), neck remnant in six (35%), aneurysm reopening in six (35%) and persistent WEB filling in one aneurysm (6%). After retreatment with the WEB, two aneurysms were again additionally treated and three aneurysms are scheduled for additional treatment. Worst results were in partially thrombosed aneurysms.
Conclusion:
The WEB device for recurrent aneurysms may be a feasible and safe option, especially in wide-necked, shallow aneurysm recurrences. Results were poor in partially thrombosed recurrent aneurysms.
Related Concept Videos
Aneurysm II: Clinical Manifestations and Diagnostic Studies
Aneurysm I: Introduction
Aneurysm IV: Nursing Management
Aneurysm III: Interprofessional Care
Nursing Clinical Information System
A Nursing Clinical Information System (NCIS) is a specialized type of healthcare information system tailored to meet the unique needs of nursing practice. It incorporates the principles of nursing informatics to streamline information management and improve the quality of care delivery.
Critical attributes of NCIS include:
Clinical Trials
There are four phases in a clinical trial. A phase one...