320-row CT transcatheter aortic valve replacement planning with a single reduced contrast media bolus injection
Daddy Mata-Mbemba1, Aissam Labani2, Soraya El Ghannudi2
1Department of Diagnostic Imaging, Hospital for Sick Children, Toronto, Ontario, Canada.
Objective:
To reduce the iodine load required for CT Transcatheter Aortic Valve Replacement (TAVR) planning on a 320-row scanner by acquiring the two CT TAVR steps (ECG-gated aortic root CTA and non-gated aorto-ilio-femoral CTA) within a single contrast media bolus injection.
Methods:
50 consecutive patients (82.6±6.9 years; 56% female) were prospectively enrolled and underwent a TAVR planning using a 320-row CT, with ECG-gated aortic root CTA immediately followed by a non-gated aorto-iliac acquisition, all within a single bolus of 40-70mL of Iohexol 350mgI/mL. The Iodine load, image quality, SNR, CNR and radiation dose were compared using a Mann-Whitney test to that of 24 consecutive patients (84.3±4.8 years, 58% female) previously imaged on a 64-row scanner with a conventional two-step protocol.
Results:
Iodine load was reduced by 44%. All examinations were of diagnostic quality, with improvement of the aortic root CTA image quality (4.9±0.3 versus 4.6±0.5, p<0.01) and a non-significant decrease of the aorto-iliac CTA image quality (4.7±0.6 versus 4.9±0.3, p = 0.07). SNR and CNR were significantly improved in the aortic root CTA (14.0±5.3 and 10.4±4.5 versus 10.3±4.2 and 6.8±3.3, p<0.01 for both) and non-significantly higher in the aorto-iliac CTA (16.5±8.0 and 14.1±7.9 versus 14.7±5.5 and 12.5±5.0, p = 0.42 and p = 0.66). Total radiation dose was reduced by 32%.
Conclusion:
320-row CT scanner enables a 44% reduction of iodine load in TAVR planning, while maintaining excellent aorto-ilio-femoral arterial enhancement and lowering radiation dose.
More Related Videos
14:14Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Nonlinear Pharmacokinetics: Drug Elimination for IV Bolus Injection
Following the administration of a single intravenous (IV) bolus injection, we can determine the concentration of the drug in the plasma at any given time. This calculation is achieved using a specific equation that integrates the values of Vmax and KM.
We can also...
Sampling Plans
Random sampling is a method where each member of the population has an equal chance of being selected for the sample. It involves selecting individuals randomly, often using random number generators or lottery-type methods. For example, when analyzing the properties of a...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...
Planning Nursing Care I