The performance of a multi guard ring (MGR) diode for clinical electron beams dosimetry
C R Nascimento1, V K Asfora2, J A C Gonçalves3
1Departamento de Energia Nuclear- UFPE, Av. Prof. Luiz Freire, 1000, 50740-545, Cidade Universitária, Recife, PE, Brazil; Instituto Federal de Educação, Ciência e Tecnologia de Pernambuco - IFPE, Recife, PE, Brazil.
Abstract:
The dosimetric response of a multi guard ring structure (MGR) diode has been studied with clinical electron beam energies from 5 MeV to 15 MeV. The results showed that the MGR dose response is linear in the range of 5-320 cGy and presents reproducibility with variation coefficients less than 0.4%. The field output factors measured with the MGR agreed within 2% with those measured with an ionization chamber. This study evidences that this diode can be used for clinical electron beam dosimetry.