Evaluation of the SD Bioline Strep A Ultra Test in Relation With Number of Colony Forming Units and Color Intensity
Sang Hyuk Ma1, Won Hee Choi2, Hyunwoong Park3
1Department of Pediatrics, Changwon Fatima Hospital, Changwon, Korea.
Background:
The SD Bioline Strep A Ultra (SD, Yongin, Korea) is a recently developed rapid antigen detection test (RADT) for diagnosing bacterial pharyngitis caused by Group A Streptococcus, We evaluated the performance of SD Bioline Strep A Ultra, using the number of colony forming units and color intensity.
Methods:
Three throat swabs each were taken from 343 children with pharyngitis who visited pediatric clinics. We evaluated the performance of SD Bioline Strep A Ultra and compared its positive rate with the number of colony forming units, using the Fisher exact test.
Results:
The sensitivity, specificity, positive predictive value, and negative predictive value (95% confidence interval) were 97.4% (94.0-99.1%), 90.8% (85.0-94.9%), 93.0% (88.5-96.1%), and 96.5% (92.0-98.9%), respectively. Positive rate significantly differed by number of colony forming units (P=0.021). ROC plot for color intensity showed 0.938 of AUC (area under curve).
Conclusions:
SD Bioline Strep A Ultra showed excellent performance, and its positive rate differed by the number of colony counts. This RADT could be used as a sensitive and semi-quantitative method detecting bacterial pharyngitis.
More Related Videos
08:31Identification and Isolation of Burst-Forming Unit and Colony-Forming Unit Erythroid Progenitors from Mouse Tissue by Flow Cytometry
Published on: November 4, 2022
12:55Fast Colony Forming Unit Counting in 96-Well Plate Format Applied to the Drosophila Microbiome
Published on: January 13, 2023
Related Concept Videos
Measurement: Standard Units
Measurement: Derived Units
Colors and Magnetism
When atoms or molecules absorb light at the proper frequency, their electrons are excited to higher-energy orbitals. For many main group atoms and molecules, the absorbed photons are in the ultraviolet range of the electromagnetic spectrum, which cannot be detected by the human eye. For coordination compounds, the energy difference between the d orbitals often allows photons in the visible range to be absorbed and emitted, which is seen as colors by the human...
Color Vision
Sound Intensity
Units of Measurement
There are various well-known historical measurement systems, such as the Babylonian system, the Roman system, the Egyptian system, the Olympian system, the British system, and the Indus...