Use of a laser-guided collimation system to perform direct kilovoltage x-ray spectra measurements on a linear
Zachary Grelewicz1, Andrew H Belcher1, Rodney D Wiersma1
1Department of Radiation and Cellular Oncology, University of Chicago, Chicago, IL, 60637-1470, USA.
Purpose:
The increased use of image-guided radiation therapy (IGRT) has led to increased use of kV on board imaging (OBI) devices. At present, directly measured OBI beam quality data have only been reported in terms of half-value layers (HVL). However, the HVL metric alone does not give the full OBI energy spectra as needed for accurate beam modeling. Although direct kV spectrometer devices exist they typically suffer from detector pile-up when used with OBI sources. We therefore present, for the first time, a novel laser-guided collimation system that allows direct measurement of the full energy spectrum for clinical OBI systems.
Methods:
Several clinically relevant spectra (80, 100, and 125 kVp), with and without the half bow-tie filter, were measured using a thermoelectric cooled cadmium telluride (CdTe) detector paired with a multichannel analyzer. To prevent detector saturation, the photon flux at the detector was reduced by use of an in-house designed laser-guided collimation system. After applying energy bin corrections, direct spectroscopic measurements were compared to Monte Carlo (MC) simulated spectra in order to verify accuracy of collected data. Both percent depth dose (PDD) curves and digitally reconstructed radiographs (DRR) were compared using the measured vs MC spectra.
Results:
Measured and MC spectra agree with RMSD between 1.96% and 3.29%. PDD curves generated from the measured and MC spectra were found to match except for in the small buildup region, with an overall match for the six beams ranging between 0.3% and 2.7% RMSD. DRRs matched well with a maximum difference in contrast of 1.1% and RMSD of 0.46% contrast for various materials in DRRs.
Conclusions:
The use of a laser-guided collimation system provided a method for quickly obtaining highly accurate kV spectrum data from OBI sources. For kV dose or DRR calculation, it was found that both spectra produced similar results.
Related Concept Videos
Direction of Acceleration Vectors
X-ray Imaging
Measuring Acceleration Due to Gravity
A simple pendulum can be described as a point mass and a string. Meanwhile, a physical pendulum is any object whose oscillations are similar to a simple pendulum, but cannot be modeled as a point mass on a string because its mass is distributed over a larger area. The behavior of a physical pendulum can be modeled using the principles of...
Emission Spectra
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...