Development of a PMMA phantom as a practical alternative for quality control of gamma knife® dosimetry
Jae Pil Chung1, Young Min Seong1, Tae Yeon Kim2
1Center for Ionizing Radiation, Division of Metrology for Quality of Life, Korea Research Institute of Standards and Science, 267 Gajeong-ro, Yuseong-gu, Daejon, 34311, Korea.
Background:
To measure the absorbed dose rate to water and penumbra of a Gamma Knife® (GK) using a polymethyl metacrylate (PMMA) phantom.
Methods:
A multi-purpose PMMA phantom was developed to measure the absorbed dose rate to water and the dose distribution of a GK. The phantom consists of a hemispherical outer phantom, one exchangeable cylindrical chamber-hosting inner phantom, and two film-hosting inner phantoms. The radius of the phantom was determined considering the electron density of the PMMA such that it corresponds to 8 g/cm2 water depth, which is the reference depth of the absorbed dose measurement of GK. The absorbed dose rate to water was measured with a PTW TN31010 chamber, and the dose distributions were measured with radiochromic films at the calibration center of a patient positioning system of a GK Perfexion. A spherical water-filled phantom with the same water equivalent depth was constructed as a reference phantom. The dose rate to water and dose distributions at the center of a circular field delimited by a 16-mm collimator were measured with the PMMA phantom at six GK Perfexion sites.
Results:
The radius of the PMMA phantom was determined to be 6.93 cm, corresponding to equivalent water depth of 8 g/cm2. The absorbed dose rate to water was measured with the PMMA phantom, the spherical water-filled phantom and a commercial solid water phantom. The measured dose rate with the PMMA phantom was 1.2% and 1.8% higher than those measured with the spherical water-filled phantom and the solid water phantom, respectively. These differences can be explained by the scattered photon contribution of PMMA off incoming 60Co gamma-rays to the dose rate. The average full width half maximum and penumbra values measured with the PMMA phantom showed reasonable agreement with two calculated values, one at the center of the PMMA phantom (LGP6.93) and other at the center of a water sphere with a radius of 8 cm (LGP8.0) given by Leksell Gamma Plan using the TMR10 algorithm.
Conclusions:
A PMMA phantom constructed in this study to measure the absorbed dose rates to water and dose distributions of a GK represents an acceptable and practical alternative for GK dosimetry considering its cost-effectiveness and ease of handling.
Related Concept Videos
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Alternative RNA Splicing
There are five types of alternative RNA splicing that vary in the ways the pre-mRNA segments are removed or retained in the mature mRNA. The first...
Alternative RNA Splicing
Null and Alternative Hypotheses
The null hypothesis, denoted by H0 is a statement of no difference between the variables—they are not related. This can often be considered the status quo. As a result if you cannot accept the null, it requires some action.
The alternative hypothesis, denoted by H1 or Ha, is a claim about the...
Quality Assurance
Quality of Water