Routine Screening Method for Microparticles in Platelet Transfusions
Published on: January 31, 2018
Comparative evaluation of platelet counts in two hematology analyzers and potential effects on prophylactic platelet
Kornelia Hummel1, Michaela Sachse1, Johannes J M L Hoffmann2
1MEDILYS Laboratories LmbH, Central Laboratory of the Asklepios Clinics, Hamburg, Germany.
Background:
Decisions on prophylactic platelet (PLT) transfusions are generally based on the recipient's PLT count, but few clinicians are aware of precision and accuracy of the PLT counting methods used by the clinical laboratory. Each PLT counting technology has its specific inaccuracy, especially in thrombocytopenic samples and therefore may impact decisions on PLT transfusions.
Study Design And Methods:
Five routine PLT counting methods available in two hematology analyzers (Sysmex XN-2000 and Abbott CELL-DYN Sapphire) were investigated (impedance and optical on both analyzers and fluorescent on XN-2000), using the CD61 immunologic PLT method as a reference. The impact of counting inaccuracy on imaginary transfusion decisions was examined at various common PLT thresholds.
Results:
In total 341 samples were analyzed, 178 of which had PLT counts of less than 35 × 109 /L. Despite excellent overall correlation with the reference method (r > 0.99), thrombocytopenic samples showed only modest correlation for impedance and XN-2000 optical methods. Sapphire optical and XN-2000 fluorescent methods correlated very well with the reference, albeit with bias in the very low range. We noticed potential risk of undertransfusion (ranging from 2% to 90%), depending on the threshold used. The risk of overtransfusion was small (<10%).
Conclusions:
The XN-2000 fluorescent PLT counting method showed excellent correlation with the CD61 reference count, closely followed by the CELL-DYN Sapphire optical method. XN-2000 impedance and optical and Sapphire impedance methods are not accurate enough for basing transfusion decisions on. Only XN-2000 fluorescent, Sapphire optical, and CD61 methods are sufficiently accurate for making appropriate clinical decisions in patients with severe thrombocytopenia.
More Related Videos
09:38A Microfluidic Flow Chamber Model for Platelet Transfusion and Hemostasis Measures Platelet Deposition and Fibrin Formation in Real-time
Published on: February 14, 2017
10:25Microfluidic Flow Chambers Using Reconstituted Blood to Model Hemostasis and Platelet Transfusion In Vitro
Published on: March 19, 2016
Related Concept Videos
Structure and Function of Platelets
Platelets are continually replenished, circulating in the bloodstream for 9-12 days before being removed by phagocytes, primarily in the spleen. A microliter of circulating blood contains between 150,000 and 450,000...
Formation of the Platelet Plug
As the injured blood vessel contracts, endothelial cells undergo contraction, revealing collagen fibers in the basement membrane and underlying connective tissue. Furthermore, the plasma membrane of endothelial cells becomes adhesive, preparing the site for platelet adhesion. Platelets...
Blood Transfusion
Blood Transfusion Overview
A blood transfusion is a medical procedure used to replace blood lost due to injury, surgery, or to treat conditions such as anemia or cancer. During a transfusion, donor blood is...
Blood Transfusion and Agglutination
History
The history of blood transfusion dates back to the 17th century, when early attempts were made in animals. In 1818 James Blundell, a British doctor, performed the first successful human blood transfusion. Later in 1900, Karl...
Framing Effects
Decision Making
Automatic decision-making is fast, intuitive, and relies on gut feelings...