Dual edge apparatus and algorithm for measurement of x-ray beam spot parameters
Andrew Jeung1, Liangjia Zhu1, Marcelo Cassese1
1Varian Medical Systems Inc., Palo Alto, CA, 94304, USA.
Purpose:
The shape, size, and location of the x-ray beam spot (where the electron beam strikes the target) in a linac-based radiation therapy machine are of potential clinical significance. Established techniques to measure the beam spot parameters involve specialized hardware and typically assess size and shape of the beam spot or its position, but not both. A simple apparatus and algorithm for measuring all beam spot parameters simultaneously is proposed here.
Methods:
The apparatus is composed of two partially transmitting edge plates mounted at different vertical positions. The mount for the apparatus slides into the accessory tray of the linac treatment head so that it rotates with the collimator, and it is imaged by the existing electronic portal imaging device (EPID) over multiple collimator angles. A software algorithm takes the acquired images and uses a parallel-beam CT reconstruction technique to compute beam spot size, shape, and position in one computation. In addition, the wobble of the collimator assembly can be estimated. The overall method was validated with both Monte Carlo simulation and with comparison to in-house spot camera measurements on a radiation therapy system.
Results:
The algorithm correctly predicted the beam spot parameters used for the Monte Carlo simulation to better than 50 μm accuracy in all cases. Furthermore, results from the dual edge method matched spot camera results with 30 μm accuracy for beam spot size and shape, with 80 μm average accuracy for beam spot position, and better than 200 μm accuracy for collimator assembly wobble.
Conclusions:
We have developed a combination dual edge apparatus and image processing algorithm that, when used on a radiotherapy linac with an EPID, can accurately determine the size and shape of the electron beam spot, its position relative to collimator rotation axis, and the wobble of the collimator assembly.
Related Concept Videos
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Golgi Apparatus
Golgi Apparatus
The Golgi apparatus is a eukaryotic organelle that has a distinctive ribbon-like appearance. It is a primary sorting and dispatch station for cargo arriving from the ER. Newly arriving vesicles enter the cis face of the Golgi, closest to the ER, and are...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Trial and Error and Algorithm
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...